Z-1170-2013Class I

The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in...

Abbott Medical / initiated 2013-03-18 / Terminated / root cause: design

Hospira Inc. began a recall of The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is... on . FDA classified it as Class I and gives the reason as "Pump shutting off during use without warning". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1170-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients. The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required.
Product code
FRN Pump, Infusion
Reason for recall
Pump shutting off during use without warning.
Root cause tag
design
FDA root cause
Device Design
Action
Hospira sent an Important Device Information letter dated March 18, 2013, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect the affected product for signs of leakage, corrosion or other damage prior to each use. In addition, each time the batteries are replaced the battery compartment should be inspected for damage. If a device exhibits damage caused by leakage, it should immediately be removed from clinical service and Hospira should be contacted at 1-800-441-4100 to arrange for return of the device for repair. If the product has been further distributed they should notify their consignees and confirm they have done so by returning the reply form to Stericycle. For questions regarding this recall call 224-212-4892.
Quantity in commerce
List Number 13086: 286 pumps; List Number 13087: 45,376 pumps; List Number 13088: 23,942 pumps
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
Product codes and lots
List Numbers: 13086, 13087, 13088

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1170-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1169-2013Class ITerminatedVoluntary: Firm initiated
Z-1170-2013Class ITerminatedVoluntary: Firm initiated