Z-1556-2013Class II
VITROS Chemistry Products-HCY Reagent packaged in a kit carton. Catalog# 6802002. Used to quantitatively...
Ortho-Clinical Diagnostics, Inc. / initiated 2013-04-01 / Terminated / root cause: manufacturing process
Ortho-Clinical Diagnostics began a recall of VITROS Chemistry Products-HCY Reagent packaged in a kit carton. Catalog# 6802002. Used to quantitatively measure total homocysteine (HCY) concentration in... on . FDA classified it as Class II and gives the reason as "The firm is recalling the specific lot due to lower than expected results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1556-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS Chemistry Products-HCY Reagent packaged in a kit carton. Catalog# 6802002. Used to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma.
- Product code
- LPS Urinary Homocystine (Nonquantitative) Test System
- Reason for recall
- The firm is recalling the specific lot due to lower than expected results.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Ortho Clinical Diagnostics sent an "Important Product Correction Notification" letter dated April 1, 2013 to all affected customers. Ortho Clinical Diagnostics recommends that their customers immediately discontinue using and discard all VITROS HCY Reagent remaining in your inventory for the affected lot. They have requested that their customers complete and return the attached Confirmation of Receipt form no later than April 12, 2013. Ortho Clinical Diagnostics requested that these notifications be forwarded to all laboratories and posted beside each VITROS System that utilizes VITROS HCY Reagent within their customers facilities. To place a product order for an alternate lot, please contact Ortho Clinical Diagnostics customer service representatives at 1-800-828-6316. The VITROS Chemistry Products HCY Reagent is used on the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System to quantitatively measure total homocysteine concentration in human serum and plasma.
- Quantity in commerce
- 478 total (339- Domestic) (139- International)
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AR, AZ, CA, CO, CT, FL, HI, ID, IL, KY, LA, ME, NJ, OK, TN, TX, WA, and WV, and the countries of Colombia, India, Japan, Mexico, Puerto Rico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, and Spain.
- Product codes and lots
- Lot # 1529-17-2636, expiration date 12 June 2013
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1556-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.