Z-1245-2013Class I

FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle...

Abbott Medical / initiated 2013-04-15 / Terminated / root cause: software

Abbott Diabetes Care, Inc. began a recall of FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc... on . FDA classified it as Class I and gives the reason as "AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1245-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
FreeStyle InsuLinx Blood Glucose Monitoring System: For In Vitro Diagnostic Use only. Use only with FreeStyle InsuLinx test strips. Abbott Diabetes Care, Inc. 1360 South Loop Road Alameda, CA 94502 USA The FreeStyle InsuLinx Blood Glucose Test Strips are for use with the FreeStyle InsuLinx Blood Glucose Meter to quantitatively measure glucose in capillary whole blood samples drawn from the fingertip.
Product code
NBW System, Test, Blood Glucose, Over The Counter
Reason for recall
AT rare, extremely high glucose levels (1024 mg/dL and above), the FreeStyle InsuLinx Blood Glucose Meter may provide an inaccurate reading.
Root cause tag
software
FDA root cause
Software design
Action
Abbott Diabetes Care sent an Urgent Product Recall letter dated April 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take one of the following options immediately to address the issue. 1) Customers can access a software update to resolve the issue at www.freestyleinsulinx.com/swupdate. 2) Customers can contact Abbott Diabetes Care Customers Service at 1-866-723-2697 to expedite return and replacement of the affected product at no charge. For questions customers were instructed to call Abbott Diabetes Care Customers Service at 1-866-723-2697. For questions regarding this recall call 510-239-2775.
Quantity in commerce
335,273 meters; 209,336 remain at firm.
Distribution
Nationwide Distribution
Product codes and lots
Part Numbers: 71142-70 71143-70 71145-70 71150-70: All serial and lot numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1245-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1245-2013Class ITerminatedVoluntary: Firm initiated