Z-1216-2013Class II

KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II...

Karl Storz Endoscopy-America, Inc. / initiated 2012-08-16 / Terminated / root cause: manufacturing process

Karl Storz Endoscopy America Inc began a recall of KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Karl Storz has confirmed that the Photodynamic Diagnosis (PDD) D-Light C System instruction manuals have important safety-related changes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1216-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Karl Storz Endoscopy-America, Inc.
Product
KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional accessory to white light cystoscopy when used in combination with the diagnostic imaging drug Cysview.
Product code
OAY Light Source System, Diagnostic Endoscopic
Reason for recall
The recall was initiated because Karl Storz has confirmed that the Photodynamic Diagnosis (PDD) D-Light C System instruction manuals have important safety-related changes.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
A Safety Alert letter, dated 8/16/12, was sent to all their customers to inform them that there is a labeling error in which important safety-related changes were made to the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System instruction manuals. The correct Ethylene Oxide (EtO) Sterilization set point for Relative Humidity should read, =70% RH. The correction applies to the PDD Camera Head instruction manual, PDD Rigid Telescopes instruction manual, and Tricam SL II instruction manual, the three component manuals indicated for use within the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System. The letter instructed the customers to place each of the three enclosed stickers, in the appropriate instruction manual. Customers are instructed to cover the erroneous information with the provided stickers as indicated. Customers with any questions regarding the proper use of the device, are instructed to contact Karl Storz's Technical Support Department at 800-421-0837, Ext. 5350, during office hours; or Ext. 7222, 24 hours/7 days a week for emergencies.
Quantity in commerce
15 units
Distribution
Nationwide distribution: USA including states of: IL, CA, WV, TX, SC, WI, NY, PA, NE, NJ, MD, OH, IA, and NC.
Product codes and lots
The device Instruction Manual recall, which includes: 1) Telescope, 2) Camera Head, and 3) Camera System.   Serial Numbers: ZZ2008 ZZ2016 ZZ2012 ZZ2019 TY2024 ZZ2013 TY2026 TY2025 TY2028 TY2021 ZZ2007 ZZ2020 TX2036 TX2033 ZZ2017, and  TX2035.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1216-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1216-2013Class IITerminatedVoluntary: Firm initiated