KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II...
Karl Storz Endoscopy America Inc began a recall of KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Karl Storz has confirmed that the Photodynamic Diagnosis (PDD) D-Light C System instruction manuals have important safety-related changes". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1216-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Karl Storz Endoscopy-America, Inc.
- Product
- KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System, Instruction Manuals: PDD Telescopes, Tricam SL II, and PDD Camera Heads. Used as an optional accessory to white light cystoscopy when used in combination with the diagnostic imaging drug Cysview.
- Product code
- OAY Light Source System, Diagnostic Endoscopic
- Reason for recall
- The recall was initiated because Karl Storz has confirmed that the Photodynamic Diagnosis (PDD) D-Light C System instruction manuals have important safety-related changes.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- A Safety Alert letter, dated 8/16/12, was sent to all their customers to inform them that there is a labeling error in which important safety-related changes were made to the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System instruction manuals. The correct Ethylene Oxide (EtO) Sterilization set point for Relative Humidity should read, =70% RH. The correction applies to the PDD Camera Head instruction manual, PDD Rigid Telescopes instruction manual, and Tricam SL II instruction manual, the three component manuals indicated for use within the KARL STORZ Photodynamic Diagnosis (PDD) D-Light C System. The letter instructed the customers to place each of the three enclosed stickers, in the appropriate instruction manual. Customers are instructed to cover the erroneous information with the provided stickers as indicated. Customers with any questions regarding the proper use of the device, are instructed to contact Karl Storz's Technical Support Department at 800-421-0837, Ext. 5350, during office hours; or Ext. 7222, 24 hours/7 days a week for emergencies.
- Quantity in commerce
- 15 units
- Distribution
- Nationwide distribution: USA including states of: IL, CA, WV, TX, SC, WI, NY, PA, NE, NJ, MD, OH, IA, and NC.
- Product codes and lots
- The device Instruction Manual recall, which includes: 1) Telescope, 2) Camera Head, and 3) Camera System. Serial Numbers: ZZ2008 ZZ2016 ZZ2012 ZZ2019 TY2024 ZZ2013 TY2026 TY2025 TY2028 TY2021 ZZ2007 ZZ2020 TX2036 TX2033 ZZ2017, and TX2035.