Z-1142-2013Class II

GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System

GE Healthcare / initiated 2011-10-21 / Terminated / root cause: labeling and UDI

GE Healthcare, LLC began a recall of GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System on . FDA classified it as Class II and gives the reason as "As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1142-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare, Innova 2100IQ, Innova 3100IQ, Innova 4100IQ, Cardiovascular and Interventional Imaging System.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
As part of GE Innova IQ table introduction, a label was designed to be affixed on the front Cover of the Gantry Pivot section. This safety label is missing on systems manufactured between March 2010 and July 2011. In the situation a staff member (doctor, nurse, operator, anesthetist, etc&) stands in the region between the C-arm gantry pivot and the head section of the table, there is a risk of co
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
A field action was initiated on October 21, 2011, and completed March 9, 2012, . The issue was corrected to all affected systems by applying the missing label onto the gantry. No recall letter was issued. For further questions please call (262) 513-4122.
Quantity in commerce
48
Distribution
Worldwide Distribution - Nationwide Distribution including the states of FL, IL, IN, LA, MI, MN, NJ, NM, NY, OH, PR, SC, TX and VA and the countries of Vietnam, Latvia, Italy, Israel, Indonesia, Hong Kong, United Kingdom, Finland, Spain, Czech Republic, Canada, Bahrain and Australia.
Product codes and lots
Serial Number 00000000321WVO 00000000353WV3 00000000000000 00000000346WV7 00000000292WW3 00000000327WV7 00000000333WV5 00000000325WV1 00000000298WV0 00000000301WV2 00000000271WV7 00000000250WV1 00000000251WV9 00000000332WV7 00000000263WV4 00000000275WV8 00000000335WV0 00000000249WV3 00000000303WV8 00000000331WV9 00000000255WV0 00000000337WV6 00000000359WV0 00000000362WV4 00000000352WV5 00000000318WV6 00000000324WV4 00000000355WV8 00000000354WV1 00000000329WV3 00000000274WV1 00000000267WV5 00000000305WV3 00000000322WV8 00000000317WV8 00000000132WV1 00000000280WV8 00000000026WV5 00000000254WV3 00000000420WV0 00000000297WV2 00000XXX299WV8 00000000257WV6 00000000357WV4 00000000253WV5 00000000339WV2 00000000265WV9 00000000252WV7

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1142-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1142-2013Class IITerminatedVoluntary: Firm initiated