Z-1706-2017Class II

Mobile Diagnostic X-Ray System with Digital Panel

Philips North America LLC / initiated 2012-11-28 / Terminated / root cause: software

Philips Healthcare Inc. began a recall of Mobile Diagnostic X-Ray System with Digital Panel on . FDA classified it as Class II and gives the reason as "When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1706-2017
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Mobile Diagnostic X-Ray System with Digital Panel
Product code
IZL System, X-Ray, Mobile
Reason for recall
When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image.
Root cause tag
software
FDA root cause
Software design
Action
Philips Medicals Systems distributed an "URGENT Field Safety Notice" alerting customers of the defect on November 28, 2012. Customers are requested to manually rotate images by using the rotation tool.Customers are also recommended to check whether the automatic marker is set correctly, especially after a manual image rotation is performed. The marker can be corrected manually with the annotation tool.
Quantity in commerce
220 systems
Distribution
Nationwide (AR, AZ, CA, DE, FL, GA, IA, IL, IN, KS, MD, MI, MN, NC, NM, NY, OH, OR, PA, SC, TX, VT, WI)
Product codes and lots
*** All units ***

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1706-2017
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1706-2017Class IITerminatedVoluntary: Firm initiated