cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from...
Roche Molecular Systems, Inc. began a recall of cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes on . FDA classified it as Class II and gives the reason as "In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1346-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Molecular Systems Inc
- Product
- cobas 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
- Product code
- OOI Real Time Nucleic Acid Amplification System
- Reason for recall
- In rare instances, channel shifted results have been generated with cobas 4800 assays run on v1.1.1 of the cobas 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Roche sent an "Urgent Medical Device Correction" (UMDC) notification with response form dated April 1, 2013, to all affected customers via UPS Ground on April 1, 2013. The letter identified the product, the problem, and the action to be taken by the customer. A contact number (Roche Diagnostic Technical Support) is provided for technical support 24 hours a day, seven days a week at phone number: 1-888-912-7090.
- Quantity in commerce
- 425 instruments
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Material Number 05200881001 - all lot/serial numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1346-2013 | Class II | Terminated | Voluntary: Firm initiated |