Z-1277-2013Class II

Olympus SurgMaster UES-40 electrosurgical unit ("UES-40")

OLYMPUS Corporation of the Americas / initiated 2009-11-10 / Terminated / root cause: software

Olympus America Inc. began a recall of Olympus SurgMaster UES-40 electrosurgical unit ("UES-40") Product Usage: Designed for use in a medical facility under the supervision of a trained physician... on . FDA classified it as Class II and gives the reason as "Olympus initiated a software upgrade and enhancement in 2009 and completed the action in 2010. The upgrade and enhancement were initiated due to complaints of overheating, smoke vapors and failure to operate associated...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1277-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus SurgMaster UES-40 electrosurgical unit ("UES-40") Product Usage: Designed for use in a medical facility under the supervision of a trained physician. It has been designed for general (open) and endoscopic surgery including urology, gynecology, respiratory and gastroenterology in conjunction with Olympus designed electrosurgical accessories, endoscopes (fiberscopes, videoscopes and rigid scopes) applicable for electrosurgery (cutting and coagulation), light sources and other ancillary equipment.
Product code
KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Reason for recall
Olympus initiated a software upgrade and enhancement in 2009 and completed the action in 2010. The upgrade and enhancement were initiated due to complaints of overheating, smoke vapors and failure to operate associated with the Olympus SurgMaster electrosurgical unit, model UES-40 ("UES-40") when used in conjunction with the Olympus WA22557C electrode. The complaint investigations revealed that
Root cause tag
software
FDA root cause
Use error
Action
Olympus America Inc sent a Field Corrective Action letter dated November 10, 2009 to all affected consignees. The letter identified the affected product, problem and actions to be taked. Consignees were directed to contact Gyrus at 888-524-7266 to schedule return of the UES-40 generator for the software upgrade. The letter also directed consignees to contact OAI Customer Service at 800-848-9024 if they needed a new copy of the UES-40 Instruction Manual. For questions call 484-896-5688.
Quantity in commerce
608
Distribution
Nationwide Distribution - USA Nationwide including the states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, and WY.
Product codes and lots
Model UES-40, all serial numbers prior to 7925334 that have not been upgraded.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1277-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1277-2013Class IITerminatedVoluntary: Firm initiated