Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to...
Stryker Howmedica Osteonics Corp. began a recall of Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the... on . FDA classified it as Class II and gives the reason as "Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1210-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
- Product code
- HTW Bit, Drill
- Reason for recall
- Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Stryker Orthopaedics sent a Urgent Medical Device Removal letter dated March 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Please inform users of this Medical Device Removal and pass this notice to all those individuals who need to be aware within your organization. 2. Return all affected products available at your location to Stryker Osteosynthesis or Contact Stryker Customer Service. 3. Complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-252-3635 . 4. Keep a copy of the completed and executed Business Reply Form for your records. Report any adverse events or product quality problems to Stryker Orthopaedics: 1-866-0R-ASSIST. (1-866-672-7747). As we strive for products that meet your expectations for quality and reliability, please do not hesitate to contact us, in case you have any further questions at (201) 831-5970..
- Quantity in commerce
- 107 units
- Distribution
- Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
- Product codes and lots
- Catalog number 703615
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1210-2013 | Class II | Terminated | Voluntary: Firm initiated |