Z-1210-2013Class II

Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to...

Howmedica Osteonics Corp / initiated 2013-03-28 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the... on . FDA classified it as Class II and gives the reason as "Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1210-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Howmedica Osteonics Corp. AxSOS System Instrument. The AxSOS Locking Plate System is designed to treat periarticular or intraarticular fractures of the Distal Femur, Proximal Humerus, Proximal Tibia, and the Distal Tibia. Per the operative technique, " A 4.3 mm Drill (REF 702743) is then used to drill the core hole for the locking screw. Using a fluoroscopy, check the correct depth of the drill, and measure the length of the screw. The Drill Sleeve should now be removed, and the correct length 5.0 mm Locking Screw is inserted using the Screwdriver T20 and Screw Holding Sleeve."
Product code
HTW Bit, Drill
Reason for recall
Stryker Orthopaedics has received a report indicating that the scale of the AxSOS Calibrated Drill Bit 4.3,,X263mm, flat , manufactured by Stryker Osteosynthesis, is incorrect. The incorrect scale could result in the surgeon selecting a screw that is 10mm longer than intended.
Root cause tag
design
FDA root cause
Device Design
Action
Stryker Orthopaedics sent a Urgent Medical Device Removal letter dated March 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Please inform users of this Medical Device Removal and pass this notice to all those individuals who need to be aware within your organization. 2. Return all affected products available at your location to Stryker Osteosynthesis or Contact Stryker Customer Service. 3. Complete and sign the enclosed Business Reply Form and fax a copy to: 1-865-252-3635 . 4. Keep a copy of the completed and executed Business Reply Form for your records. Report any adverse events or product quality problems to Stryker Orthopaedics: 1-866-0R-ASSIST. (1-866-672-7747). As we strive for products that meet your expectations for quality and reliability, please do not hesitate to contact us, in case you have any further questions at (201) 831-5970..
Quantity in commerce
107 units
Distribution
Nationwide Distribution including the states of AL, AZ, IL, KY, LA, NJ, NC, FL, CA, NM, MO, TN, VA, CO, MA, TX, IN and MT.
Product codes and lots
Catalog number 703615

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1210-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1210-2013Class IITerminatedVoluntary: Firm initiated