Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1958-2013Class II

Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B...

SPINEFRONTIER, Inc. dba LESspine Innovations / initiated 2013-04-22 / Terminated / root cause: labeling and UDI

SpineFrontier, Inc. began a recall of Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A... on . FDA classified it as Class II and gives the reason as "Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1958-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SPINEFRONTIER, Inc. dba LESspine Innovations
Product
Case Lid and Base (Inspan, Invue, SLIFT) Caddy Lid (Inspan, Arena-C); Inspan Lid: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A; Invue Lid: SI70095A; S-Lift Lid: SI50070A; S-Lift Base: SI50067A; Arena-C Lid: 11-72013D, 11-72014D; Generics Lid: 12-99005A. System cases are designed to hold and secure all instrumentation to be used with a surgical system.
Product code
FSM Tray, Surgical, Instrument
Reason for recall
Non-toxic inks, used on the silk screen graphics of the case bases and lids, and caddy lids, were found to be either delaminating or running off from the substrate after repeated washing and sterilization cycles.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 2/15/2013, SpineFrontier emailed distributors informing them that they needed specific parts returned to work towards completing the conversion of product graphics.
Quantity in commerce
258 units (30 in distribution at the time of the recall)
Distribution
Worldwide Distribution-USA including the states of FL, OH, MA, and TX, and the countries of Mexico, and Dominican Republic.
Product codes and lots
Inspan Lid: Part Number: 12-60039A, 11-60042B, 11-60043B, 11-60045B, 11-60046C, 11-60045A, 11-60046A, 11-60053A, Lot Numbers: 7197.024, 011649, 011498; Invue Lid: SI70095A, Lot Number: 6250.009-R; S-Lift Lid: SI50070A, Lot Number: 7201.008; S-Lift Base: SI50067A; Lot Numbers: 7183.007R, 7201.007; Arena-C Lid: 11-72013D, 11-72014D, Lot Numbers: 011969-001, 011969-002; Generics Lid: 12-99005A, Lot Number: 8372.001.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1958-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1958-2013Class IITerminatedVoluntary: Firm initiated