Z-1869-2013Class II

Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized...

SPINEFRONTIER, Inc. dba LESspine Innovations / initiated 2012-10-05 / Terminated / root cause: design

SpineFrontier, Inc. began a recall of Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion within the case, 8mm and 9mm pedicle screws... on . FDA classified it as Class II and gives the reason as "Two sections that are designated for the 8mm and 9mm screws were switched; the holes that hold the 8mm screws are designed for 9mm screws, and the holes that hold the 9mm screws are designated for 8mm screws". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1869-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SPINEFRONTIER, Inc. dba LESspine Innovations
Product
Caddy - PedFuse, Screw Caddy, 8MM and 9MM, Part Number: 11-80009 Rev. B The caddy holds, in an organized fashion within the case, 8mm and 9mm pedicle screws used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients
Product code
FSM Tray, Surgical, Instrument
Reason for recall
Two sections that are designated for the 8mm and 9mm screws were switched; the holes that hold the 8mm screws are designed for 9mm screws, and the holes that hold the 9mm screws are designated for 8mm screws.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, SpineFrontier, sent an Advisory Notice entitled "URGENT May 1, 2013, PedFuse Screw Caddy Removal" dated May 1, 2013 to their consignees/customers. The notice described the product, problem and action to be taken. Consignees/customers were informed that a removal has been initiated to remove the parts from the field to ensure that all sets contain correctly marked screw caddies. Consignees/customers wre instructed to complete and return the URGENT PedFuse Screw Caddy Removal ACKNOWLEDGEMENT FORM via fax to: SpineFrontier, Inc., 978-232-3991 or email: [email protected]. Replacement caddies will be shipped upon receipt of completed acknowledgement form and RMA request for part return. Please feel free to contact SpineFrontier at 978-232-3990 with any questions.
Quantity in commerce
11 caddies
Distribution
US distribution in states of: CO, DC, MA and TX.
Product codes and lots
31176, 018652, 018645R, 42011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1869-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1869-2013Class IITerminatedVoluntary: Firm initiated