KRD1 PedFuse Rods, Model: IM80100-13 Rev B
SpineFrontier, Inc. began a recall of KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod... on . FDA classified it as Class II and gives the reason as "One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1737-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SPINEFRONTIER, Inc. dba LESspine Innovations
- Product
- KRD1 PedFuse Rods, Model: IM80100-13 Rev B Product Usage: Pedicle rods are cylindrical rods which act as a linking mechanism between pedicle screws. The rod sits with the pedicle screw tulip and they are locked into place with a set screw or locking cap. The rod may be placed along several pedicle screws. Component rods are used during implantation of pedicle screw system devices intended to provide immobilization and stabilization of spinal segments of skeletally mature patients.
- Product code
- MNI Orthosis, Spinal Pedicle Fixation
- Reason for recall
- One of the rods in each of two test constructs did not meet fatigue requirement as outlined in ASTM F1717-04 dynamic compression testing 5 million cycles.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- SpineFrontier issued a Recall Notification to the affected distributor via telephone. The distributor was instructed to return the affected product for a replacement product.
- Quantity in commerce
- 2 rods
- Distribution
- USA Nationwide Distribution including the state of TX
- Product codes and lots
- Lot Number: 002876-007R
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1737-2013 | Class II | Terminated | Voluntary: Firm initiated |