Z-1957-2013Class II

Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis...

SPINEFRONTIER, Inc. dba LESspine Innovations / initiated 2012-05-09 / Terminated / root cause: design

SpineFrontier, Inc. began a recall of Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal... on . FDA classified it as Class II and gives the reason as "The nipple of the tipped driver broke off when surgeons were attempting to insert the screw". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1957-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SPINEFRONTIER, Inc. dba LESspine Innovations
Product
Invue Solid Driver, Tipped, AO. Part Number: 11-70083. Allows the user to retain the screw when changing axis of insertion during fixation of spinal intervertebral.
Product code
HXX Screwdriver
Reason for recall
The nipple of the tipped driver broke off when surgeons were attempting to insert the screw.
Root cause tag
design
FDA root cause
Device Design
Action
On May 9, 2012 an email was sent to consignees requesting that the product be returned for servicing and that a replacement part would be sent.
Quantity in commerce
9
Distribution
Distributed in DC and the states of MA, and OK.
Product codes and lots
Lot Number: 011836-0001, 011981-001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1957-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1957-2013Class IITerminatedVoluntary: Firm initiated