The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment...
Alcon Grieshaber AG began a recall of The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery on . FDA classified it as Class II and gives the reason as "Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1506-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Grieshaber AG
- Product
- The Constellation Vision System Pneumatic Handpiece. Indicated for both anterior and posterior segment opthalmic surgery.
- Product code
- HQE Instrument, Vitreous Aspiration And Cutting, Ac-Powered
- Reason for recall
- Alcon is conducting a Medical Device Removal for the Constellation Pneumatic Handpiece, DSP, product number 725.01. This action is being taken because the forceps or scissors tips (sold separately) may dislodge from the hand piece, if not attached by the user according to the Directions for Use.
- FDA root cause
- Use error
- Action
- Alcon sent an "URGENT: MEDICAL DEVICE REMOVAL" letter on April 17, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact your local Alcon Sales Representative for questions regarding this notice.
- Quantity in commerce
- 8,455 units
- Distribution
- Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Canada, Chile, China, Egypt, France, Germany, Great Britain, India, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Kyrgyzstan, Lebanon, Korea, Malaysia, Mexico, Pakistan, Peru, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Turkistan, Venezuela, and Veit Nam.
- Product codes and lots
- All Lots of the Constellation Vision System Pneumatic Handpiece, DSP, product number 725.01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1506-2013 | Class II | Terminated | Voluntary: Firm initiated |