Z-0615-2014Class II

Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated...

Synthes GmbH / initiated 2013-04-15 / Terminated / root cause: manufacturing process

Synthes USA HQ, Inc. began a recall of Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement... on . FDA classified it as Class II and gives the reason as "Synthes received a complaint from a customer in which the Hand Switch for the Electric Pen Drive was not labeled correctly. The "on" and "lock" labels were interchanged". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0615-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated for screw insertion, pin and wire replacement, cutting of bone and metal, drilling, decorticating, shaping, and smoothing of bones and teeth in a wide variety of surgical procedures, including, but not limited to general orthopaedic trauma, foot, hand, maxillofacial, neurosurgical, oral, otolaryngolical, reconstructive, and spine surgery.
Product code
HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
Reason for recall
Synthes received a complaint from a customer in which the Hand Switch for the Electric Pen Drive was not labeled correctly. The "on" and "lock" labels were interchanged.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Synthes sent an Urgent Notice: Medical Device Recall letter dated April 15, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to examine their inventory for the affected product, remove from inventory, and return them to Synthes immediately for replacement. Customers were instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization (RA) Number. Complete the Verification Section at the end of the letter by checking the appropriate box indicating affected product has been located. Return the Verification Form with the product to: Credit/Returns, Synthes 1101 Synthes Avenue Monument, CO 80132 If customers do not have the affected product they should still complete the Verifcation Section at the end of the letter by checking the appropriate box and faxing the document to 610-251-9005 or email to: [email protected] Customers with questions were instructed to call 610-719-5450 or email [email protected]. For questions regarding this recall call 610-719-5000.
Quantity in commerce
290
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Mexico and Canada. .
Product codes and lots
Part Number 05.001.012, lot number AV18166

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0615-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0615-2014Class IITerminatedVoluntary: Firm initiated