Z-2278-2013Class II

Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for...

Synthes GmbH / initiated 2013-03-08 / Terminated / root cause: labeling and UDI

Synthes USA HQ, Inc. began a recall of Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus... on . FDA classified it as Class II and gives the reason as "Recall is being initiated due to the part being mislabeled (Part Number 241.267 was mislabeled with Part Number 241.276)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2278-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes 3.5 MM LCP(R) Distal Humerus System The Synthes 3.5 MM LCP Distal Humerus System is intended for fixation of fractures of the distal humerus, olecranon, and ulna in adults and adolescents (12-21) in which the growth plates have fused. Specifically, distal humerus plates are indicated for intra-articular fractures, comminuted supracondylar fractures, osteotomies, malunions and non-unions of the distal humerus.
Product code
KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
Reason for recall
Recall is being initiated due to the part being mislabeled (Part Number 241.267 was mislabeled with Part Number 241.276).
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Synthes sent an Urgent Notice Medical Device Recall letter dated March 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to obtain a Return Authorization Number, complete the Verification Section at the end of the letter by the checking the appropriate box indicating affected product located. Return the Verification Form with the product to: Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132. Customers with questions were instructed to call 610-719-5450 or email [email protected]. For questions regarding this recall call 610-719-5000.
Quantity in commerce
2
Distribution
Nationwide Distribution including CA,and VA. .
Product codes and lots
Part Number 241.267, lot number 8037923

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2278-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2278-2013Class IITerminatedVoluntary: Firm initiated