Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250...
Philips Medical Systems (Cleveland) Inc began a recall of Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000... on . FDA classified it as Class II and gives the reason as "Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1338-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Brilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250; Brilliance CT 6-Slice Water, System Code: 728255; MX8000 IDT 16, System Code: 728120 & MX8000 IDT 10, System Code: 728125 Computed Tomography X-ray systems. Distributed by Philips Medical Systems (Cleveland), Inc., 595 Miner Road, Cleveland, OH 44143 Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips Healthcare received a report from the field stating when they viewed the CTDIVol information from the PACS system, the calculations were too high. Instead of storing the CTDIVol value for each slice in the DICOM tag, the system took the sum of all CTDIVol values from all the images in the dataset and stored this value in the DICOM tag, resulting in incorrect information.
- FDA root cause
- Use error
- Action
- The firm, Philips Healthcare, sent an "Urgent - Medical Device Correction" letter dated March 15, 2013, to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to disregard unusually higher CTDIvol value for Type-3 DICOM tag 0018:9345 in the PACS system, if seen. The firm is implementing software update versions 4.2.2 or 2.2.2 to correct the issue. A Philips Field Service Engineer will contact you to schedule the software installation at your site. If you need any further information or support concerning this issue, please contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customers Care Solutions Center (1-800-722-9377, Option 5: Enter Site ID or follow the prompts).
- Quantity in commerce
- 515 units
- Distribution
- Worldwide distribution: USA (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, VA, VT, WA, WI, WV and WY; countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Denmark, France, Germany, Gibraltar, India, Iraq, Ireland, Israel, Italy, Japan, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Philippines, Poland, Portugal, Singapore, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela.
- Product codes and lots
- System Code #728255, Serial #: 9702, 9703, 9704, 9705, 9706, 9707, 9709, 9710, 9713, 9714, 9717, 9718, 9719, 9721, 9722, 9723, 9724, 9725, 9727, 9730, 9731, 9732, 9733, 9734, 9735, 9738, 9739, 9740, 9741, 9742, 9743, 9746, 9747, 9749, 9752, 9753, 9756, 9758, 9759, 9760, 9761, 9762, 9765, 9766, 9768, 9769, 9770, 9772, 9774, 9778, 9780, 9781, 9783, 9784, 9786, 9787, 9790, 9791, 9792, 9793, 9796, 9797, 9801, 9805, 9807, 9808 & 9810. System Code #728250, Serial #: 2556, 2557, 2560, 2562, 2563, 2565, 2567, 2568, 2569, 2570, 2571, 2574, 2576, 2577, 2579, 2580, 2581, 2584, 2585, 2587, 2588, 2590, 2591, 2592, 2593, 2594, 2595, 2597, 9738 & 9741. System Code #728245, Serial #: 2562, 2564, 2567, 2573, 2576, 2582, 2584, 2585, 2586, 2587, 2589, 2590, 2594, 2596, 2597, 3014, 3091, 3105, 3151, 3217, 3235, 3287, 3303, 3312, 3339, 3340, 3341, 3342, 3343, 3344, 3346, 3348, 3351, 3353, 3355, 3356, 3357, 3358, 3359, 3360, 3363, 3364, 3366, 3369, 3371, 3373, 3374, 3377, 3378, 3379, 3380, 3381, 3382, 3383, 3385, 3388, 3392, 3393, 3394, 3397, 3400, 3404, 3406, 3407, 3408, 3410, 3413, 3414, 3415, 3416, 3419, 3423, 3425, 3427, 3429, 3431, 3432, 3433, 3435, 3436, 3437, 3438, 3439, 3440, 3441, 3443, 3445, 3446, 3448, 3449, 3451, 3452, 3453, 3454, 3455, 3458, 3459, 3461, 3462, 3464, 3468, 3469, 3470, 3471, 3473, 3474, 3475, 3477, 3478, 3479, 3481, 3482, 3483, 3484, 3485, 3486, 3488, 3489, 3491, 3492, 3493, 3495, 3496, 3497, 3498, 3499, 3500, 3501, 3502, 3503, 3504, 3505, 3508, 3509, 3510, 3511, 3512, 3513, 3514, 3515, 3517, 9716, 9718, 9719, 9724, 9732, 9741, 9743, 9746, 9749, 9778, 9780, 9787, 9790, 9797, 9801, 9808 & 9810. System Code #728120, Serial #: 2502, 2510, 2513, 2516, 2517, 2518, 2525, 2528, 2531, 2533, 2535, 2539, 2546, 3006, 3008, 3013, 3014, 3021, 3025, 3026, 3027, 3028, 3029, 3030, 3035, 3037, 3039, 3040, 3043, 3044, 3047, 3049, 3050, 3051, 3053, 3054, 3056, 3059, 3061, 3062, 3063, 3064, 3068, 3069, 3070, 3072, 3073, 3075, 3076, 3077, 3078, 3079, 3081, 3082, 3084, 3085, 3087, 3088, 3089, 3091, 3093, 3094, 3095, 3097, 3099, 3100, 3103, 3104, 3105, 3106, 3107, 3108, 3110, 3111, 3112, 3113, 3114, 3116, 3117, 3118, 3119, 3120, 3121, 3122, 3124, 3125, 3126, 3127, 3129, 3130, 3131, 3132, 3135, 3136, 3137, 3138, 3141, 3143, 3144, 3146, 3148, 3149, 3150, 3151, 3152, 3153, 3154, 3155, 3156, 3157, 3159, 3160, 3161, 3162, 3163, 3164, 3165, 3167, 3168, 3169, 3171, 3172, 3174, 3176, 3177, 3178, 3179, 3180, 3182, 3183, 3185, 3187, 3188, 3189, 3190, 3192, 3194, 3195, 3196, 3198, 3199, 3202, 3203, 3204, 3205, 3206, 3207, 3208, 3209, 3211, 3212, 3213, 3214, 3216, 3217, 3218, 3219, 3220, 3221, 3222, 3223, 3224, 3225, 3229, 3230, 3232, 3233, 3234, 3235, 3236, 3237, 3238, 3239, 3240, 3245, 3246, 3247, 3248, 3250, 3251, 3252, 3253, 3254, 3255, 3258, 3261, 3262, 3264, 3265, 3266, 3268, 3269, 3270, 3271, 3273, 3274, 3277, 3279, 3280, 3281, 3283, 3284, 3285, 3287, 3289, 3290, 3293, 3294, 3295, 3296, 3297, 3298, 3299, 3300, 3302, 3303, 3304, 3305, 3307, 3308, 3310, 3311, 3312, 3315, 3316, 3318, 3319, 3320, 3321, 3322, 3324, 3328, 3329, 3330, 3367 & 9760. System Code #728125, Serial #: 2501, 2505, 2508, 2509, 2513, 2515, 2516, 2522, 2523, 2524, 2525, 2526, 2529, 2530, 2531, 2533, 2534, 2535, 2539, 2540, 2542, 2543, 2548 & 2549.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1338-2013 | Class II | Terminated | Voluntary: Firm initiated |