Z-1376-2013Class II

NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in...

Zimmer GmbH / initiated 2013-03-27 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia on . FDA classified it as Class II and gives the reason as "Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1376-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
NEXGEN Complete Knee Solution Monoblock Tibial Drill with stop 10.7 mm. Used to prepare cylindrical holes in the native tibia.
Product code
HTW Bit, Drill
Reason for recall
Zimmer is initiating a lot specific recall of the Trabecular Metal" Tibia Stop Drill due to the potential of the outside diameter being oversized. As a result, there is a potential for the drill to not pass through the applicable drill guide.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer sent an "URGENT MEDICAL DEVICE RECALL LETTER" dated April 19, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.
Quantity in commerce
69 units
Distribution
Nationwide Distribution.
Product codes and lots
Part No. 00-5887-052-00; lots 62181290 and 62183430

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1376-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1376-2013Class IITerminatedVoluntary: Firm initiated