Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2292-2013Class II

Solitaire-35 Peek Solitaire Anterior Spinal System. The Solitaire-35 is an intervertebral body fusion device

Ebi, LLC / initiated 2013-04-09 / Terminated / root cause: manufacturing process

Ebi, Llc began a recall of Solitaire-35 Peek Solitaire Anterior Spinal System. The Solitaire-35 is an intervertebral body fusion device on . FDA classified it as Class II and gives the reason as "Several complaints were recorded regarding one of the three screws driving through the cage of the Solitaire 35 implant". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2292-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ebi, LLC
Product
Solitaire-35 Peek Solitaire Anterior Spinal System. The Solitaire-35 is an intervertebral body fusion device.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Several complaints were recorded regarding one of the three screws driving through the cage of the Solitaire 35 implant.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Biomet LLC sent "Urgent Medical Device Recall" letters/fax back response forms on Tuesday April 9, 2013 via Fed Ex to all distributors. The notification identified the problem, product involved, and risk factors. In addition, the notification provides instructions to the representatives on how to return the product to EBI, LLC. Surgeons that have implanted the product were contacted on April 12, 2013.
Quantity in commerce
1730 units
Distribution
Nationwide distribution: US including states of: CA, GA, IN, KY, MO, NY, OR, TX, and WI.
Product codes and lots
14-535140 12 mm 6-deg narrow 2327561 October-22 2302271 April-22  14-535141 14mm 6-deg narrow 2327661 November-22 2302281 April-22  14-535142  16mm 6-deg narrow 2327691 December-22 2302291 April-22  14-535143 18 mm 6-deg narrow  2327701 December-22 2302301 April-22  14-535144  20mm 6-deg narrow 2327711 December-22 2302301 April-22  14-535144 20 mm 6-deg narrow 2302311 June-22  14-535150  12mm 12-deg narrow 2327711 December-22 2302321 April-22  14-535151 14mm 12-deg narrow 2327601 October-22 2302332 April-22   14-535152 16mm12-deg narrow  2327671 November-22 2302341 April-22  14-535153 18 mm 12-deg narrow  2327751 December-22 2302351 April-22  14-535154 20 mm 12-deg narrow 2327571 December-22 2327571 November-22 2302361 April-22  14-535170 12mm 6-deg medium  2327631 October-22 2302371 April-22  14-5351711  14 mm 6-deg medium 2327641 October-22 2302381 April-22  14-535172 16 mm 6-deg medium 2327761 December-22 2302391 April-22  14-535173  18 mm 6-deg medium  2327771 Dec 22 2302401 April 22  14-535174  20 mm 6-deg medium 2302411 April-22  14-535180 12mm 12-deg medium 2327621 October-22 2302421 April-22  14-535180 12 mm 12-deg medium 2327621 October-22 2302421 April-22  14-535182 16mm 12-deg medium 2327651 October 22 2302441 April 22  14-535183  18 mm 12-deg medium  2302451 April-22  14-535184 20 mm 12-deg medium  2327581 October-22 2302461 April-22  14-535200 12 mm 6-deg wide 2302471 April-22  14-535201  14mm 6-deg wide 2327721 December-22 2302481 April-22  14-535202  16 mm 6-deg wide  2327591 October-22 2302491 March-22  14-535203 18 mm 6-deg wide 2328311 October-22 2302501 April-12  14-535204  20 mm 6-deg wide 2302511 April-22   14-535210 12 mm 12 deg wide 2302521 April 22  14-535211 14 mm 12-deg wide 2327681 Decembr-22 2302531 April-22  14-535212 16 mm 12-deg wide 2327731 December-22 2302541 April-22  14-535213 18mm 12-deg wide 2327741 December-22 2302551 May-22

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2292-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2292-2013Class IITerminatedVoluntary: Firm initiated