BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast...
Bard Peripheral Vascular Inc began a recall of BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical... on . FDA classified it as Class II and gives the reason as "Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1524-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy.
- Product code
- FZP Clip, Implantable
- Reason for recall
- Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Bard Peripheral Vascular, Inc. initiated a recall of certain product code and lot number combinations of Bard UltraClip Dual Trigger Tissue Marker by contacting customers on 04/16/2013. On 04/18/2013, BPV sent a Customer Recall notification Letter via FedEx with proof of delivery notice. The letter included reason for recall, a table of affected product code and lot numbers, instructions (Do not use or further distribute any affected product), a response form (titled Recall and Effectiveness Check Form) enclosed, and contact information.
- Quantity in commerce
- 3,085
- Distribution
- Worldwide Distribution: USA Nationwide and Internationally to Europe.
- Product codes and lots
- Affected Product Lots with the following Product Code and Lot Number combinations. Product Code 864017D 864017D 864017D 864017D 864017D 864017D 864017D 864017DL Lot Number HUWI1522 HUWI1941 HUWI1942 HUWI1943 HUWJ1851 HUWJ1852 HUWK0350 HUWJ1610
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1524-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY