Z-1524-2013Class II

BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast...

Bard Peripheral Vascular Inc / initiated 2013-04-16 / Terminated / root cause: manufacturing process

Bard Peripheral Vascular Inc began a recall of BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical... on . FDA classified it as Class II and gives the reason as "Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1524-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD Biopsy Systems UltraClip Dual Trigger breast tissue marker. Intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy.
Product code
FZP Clip, Implantable
Reason for recall
Bard Peripheral Vascular (BPV) has confirmed that some devices with the recalled product code / lot number combination may have the metal tissue marker separated from the needle prior to use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Bard Peripheral Vascular, Inc. initiated a recall of certain product code and lot number combinations of Bard UltraClip Dual Trigger Tissue Marker by contacting customers on 04/16/2013. On 04/18/2013, BPV sent a Customer Recall notification Letter via FedEx with proof of delivery notice. The letter included reason for recall, a table of affected product code and lot numbers, instructions (Do not use or further distribute any affected product), a response form (titled Recall and Effectiveness Check Form) enclosed, and contact information.
Quantity in commerce
3,085
Distribution
Worldwide Distribution: USA Nationwide and Internationally to Europe.
Product codes and lots
Affected Product Lots with the following Product Code and Lot Number combinations.   Product Code  864017D  864017D  864017D  864017D  864017D  864017D  864017D  864017DL   Lot Number   HUWI1522   HUWI1941   HUWI1942   HUWI1943   HUWJ1851   HUWJ1852   HUWK0350   HUWJ1610

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1524-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1524-2013Class IITerminatedVoluntary: Firm initiated