Z-1401-2013Class II

Alere" hCG Combo Cassette (20/10 mIU/mL) Kit

Alere San Diego, Inc. / initiated 2013-04-16 / Terminated / root cause: packaging

Alere San Diego, Inc. began a recall of Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the... on . FDA classified it as Class II and gives the reason as "The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1401-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alere San Diego, Inc.
Product
Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine, as an aid in the early detection of pregnancy. It is for healthcare professionals only.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for
Root cause tag
packaging
FDA root cause
Packaging
Action
Alere initiated this recall by sending recall notification letters, titled "URGENT MEDICAL DEVICE RECALL", dated April 16, 2013 via via fax, email or direct mail. The letter informed customers of the product recalled, reason for recall, customer required action, and contact information, "Alere Technical Services at 866-216-0094 or by e-mail at [email protected].". A verification form was attached to the notification letter.
Quantity in commerce
3975 kits
Distribution
United States Nationwide Distribution
Product codes and lots
Lot/Unit Codes: hCG1110133 & hCG1110135  Part Number: 92215  Expiration date, or Expected shelf life: 9/30/13 & 10/31/13

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1401-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1401-2013Class IITerminatedVoluntary: Firm initiated