Alere" hCG Combo Cassette (20/10 mIU/mL) Kit
Alere San Diego, Inc. began a recall of Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the... on . FDA classified it as Class II and gives the reason as "The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1401-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Alere" hCG Combo Cassette (20/10 mIU/mL) Kit. Product Usage: The AlereTM hCG Combo (20/10 mIU/mL) test is a rapid chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine, as an aid in the early detection of pregnancy. It is for healthcare professionals only.
- Product code
- JHI Visual, Pregnancy Hcg, Prescription Use
- Reason for recall
- The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Alere initiated this recall by sending recall notification letters, titled "URGENT MEDICAL DEVICE RECALL", dated April 16, 2013 via via fax, email or direct mail. The letter informed customers of the product recalled, reason for recall, customer required action, and contact information, "Alere Technical Services at 866-216-0094 or by e-mail at [email protected].". A verification form was attached to the notification letter.
- Quantity in commerce
- 3975 kits
- Distribution
- United States Nationwide Distribution
- Product codes and lots
- Lot/Unit Codes: hCG1110133 & hCG1110135 Part Number: 92215 Expiration date, or Expected shelf life: 9/30/13 & 10/31/13
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1401-2013 | Class II | Terminated | Voluntary: Firm initiated |