Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA
Canon Inc. began a recall of Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the... on . FDA classified it as Class II and gives the reason as "Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1404-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Inc.
- Product
- Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA. Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye.
- Product code
- HKO Refractometer, Ophthalmic
- Reason for recall
- Canon Inc. had received 17 claims from our foreign sales companies between October to November 2012 that measurement value of RK-F2s refractive power (SPH value) sometimes indicated abnormal value which occurs approximately 5% in probability.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Canon initiated recall by sending letters on November 22, 2012 to all affected customers. The letter identified the affected product, problem, actions to be taken and contact information.
- Quantity in commerce
- 384 units
- Distribution
- Worldwide Distribution - USA Nationwide and countries of: Australia, Japan, Korea, Singapore and United Kingdom.
- Product codes and lots
- Catalog Number 6937B002AA. All serial numbers from the first unit onward.