Z-1693-2013Class II

Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2013-05-24 / Terminated / root cause: manufacturing process

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models... on . FDA classified it as Class II and gives the reason as "Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1693-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius Service Replacement 0-Ring. Used in as a service part in 4 sub-assemblies manufactured and/or refurbished Fresenius hemodialysis machines, models 2008K, 2008K2, 2008K@Home and 2008T
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Incorrect rubber nitrile O-rings distributed instead of the correct material EPDM. (Ethylene Propylene Dine Monomer (M-class) rubber)
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Fresenius notified customers on May 24, 2013, by telephone and by certified mail with signature confirmation and fax-back form. Customers were instructed to examine their stock immediately to determine whether they have any of the affected parts on hand. If customers have the affected parts, they will be instructed to immediately discontinue use and place all 0-rings in a secure area for return to Fresenius Medical Care North America. Please contact your Fresenius Medical Care Technical Service Team at 1-800-227-2572, (choose option 4, then option 2) for instructions how to return the recalled product.
Quantity in commerce
725 units
Distribution
Nationwide Distribution including the states of AR, CA, FL, GA, KY, LA, MA, MD, NC, NJ, NY, PA, PR, SC, TX and VA.
Product codes and lots
Part Number: 640919

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1693-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1693-2013Class IITerminatedVoluntary: Firm initiated