510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1339-2013Class II

LATITUDE Patient Management System, Model 6488, Version 7.3. The LATITUDE Patient Management system is...

Boston Scientific Corporation / initiated 2013-04-10 / Terminated / root cause: software

Boston Scientific CRM Corp began a recall of LATITUDE Patient Management System, Model 6488, Version 7.3. The LATITUDE Patient Management system is intended for use to remotely communicate with a... on . FDA classified it as Class II and gives the reason as "Boston Scientific CRM is conducting a recall on the LATITUDE Patient Management System Model 6488 Version 7.3 because between March 4 and 27, 2013, a Siebel release caused ~5000 distributed PMR clinic faxes to display...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1339-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
LATITUDE Patient Management System, Model 6488, Version 7.3. The LATITUDE Patient Management system is intended for use to remotely communicate with a compatible Boston Scientific pulse generator and transfer data to a central database.
Product code
LWS Implantable Cardioverter Defibrillator (Non-Crt)
Reason for recall
Boston Scientific CRM is conducting a recall on the LATITUDE Patient Management System Model 6488 Version 7.3 because between March 4 and 27, 2013, a Siebel release caused ~5000 distributed PMR clinic faxes to display with illegible characters and not the required, Patient non-compliance information that was intended.
Root cause tag
software
FDA root cause
Software change control
Action
Boston Scientific CRM Corp began re-faxing the 4,912 previously illegible PMR notifications on April 10, 2013 and completed on April 15, 2013, successfully fulfilling labeling expectations that these notifications be communicated to the clinic. For questions regarding this recall call 651-582-4000.
Quantity in commerce
1
Distribution
Nationwide Distribution including DC and PR
Product codes and lots
n/a

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1339-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1339-2013Class IITerminatedVoluntary: Firm initiated