Z-1430-2013Class II

AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum...

Siemens Healthcare Diagnostics Inc. / initiated 2013-04-26 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood on . FDA classified it as Class II and gives the reason as "Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1430-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
AU2700 Clinical Chemistry Analyzer AU5400 Clinical Chemistry Analyzer Perform automated analysis of serum, urine, and other body fluids, including whole blood.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Beckman Coulter is recalling AU2700/AU2700 Plus & AU5400 Clinical Chemistry Analyzers becuase it was determined that the tubing/joint connections associated with these MODs may leak concentrated detergent. Beckman Coulter has received complaints that the joint connecting the tubing appears to disintegrate over time, causing a leak.
Root cause tag
design
FDA root cause
Component design/selection
Action
Beckman Coulter sent an Urgent Product Correction letter on April 26, 2013 along with a response form. The letter notified customers of the product, problem, and actions to be taken by the customers. Contact the Call Center at 800-854-3633 for questions regarding this notice.
Quantity in commerce
108 units
Distribution
Worldwide Distribution-USA Nationwide and countries of: AU, CA, DE, ES, GB, IT, PT, and JP.
Product codes and lots
All Serial Numbers  Model Numbers: AU2700, AU27000ISE, AU5421, AU5421ISE, AU5431.  Part Numbers: E0750053 N3663400  N3663700  N3664000  N3664200  N3664400  N3664600  N3664800  N3878500  1465600  64270G  E61DAU2700 N3663500  N3663600  N3663800  N3663900  N3664100  N3664300  N3664500  N3664700  N3664900  N3878600  O9102700 N3665000  N3665100  N3665200  N3665300  N3665400  N3878300  64542  A94504  N1683100  AU5431-U1  N3665500  N3665600  N3665700  N3665800  N3665900  N3878400

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1430-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1430-2013Class IITerminatedVoluntary: Firm initiated