Z-1971-2013Class II

epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2...

Epocal Inc / initiated 2013-03-13 / Terminated / root cause: manufacturing process

Epocal began a recall of epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate... on . FDA classified it as Class II and gives the reason as "This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1971-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Epocal Inc
Product
epoc BGEM Test Card (Catalogue number CT-1004-00-00). quantitative testing of samples for pH, pCO2, pO2, Sodium Potassium , Ionized Calcium, Glucose, Lactate and Hematocrit.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
This recall is being initiated because the affected lots of this product may yield results that are not consistent with stated precision performance criteria at concentration values at the high-end of the Glucose test sensor (>400 mg/dl). Specifically, after a period of storage of three (3) to four (4) months, the Glucose sensor may report low results at the high end of the measurement range.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On March 13th 2013, Epocal placed all CT-1004-00-00 test card on shipment hold. Affected product in Epocal control has been segregated and quarantined. Beginning on March 26 2013, Epocal Inc. plans to begin distribution of the Recall Notification Letter to the 4 direct consignees (customers/distributors) via certified mail. Included with the Recall Notification Letter is a faxback verification form. Notification will include all distribution channels to the end-user level. Direct consignees are instructed to discontinue use and discard all stock on hand, complete the provided faxback verification form to document disposal and contact their consignees to advise them of the notification and disposal of affected product.
Quantity in commerce
828,750 Test Card
Distribution
Worldwide Distribution-USA (nationwide) including the District of Columbia, US Virgin Islands and Puerto Rico, and the states of CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MD, MI, MN, MS, NC, ND, NE, NH, NJ, NM, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VE, WA, WI, and WV and the coutries of India, Argentina, Australia, Azerbaijan, Belarus, Canada, China, Colombia, Egypt, Germany, Guatemala, Israel, Japan, Malaysia, Mexico, Nepal, Netherlands, Peru, Philippines, Russia Fed., Singapore, South Africa, South korea, Thailand, and Vietnam.
Product codes and lots
LOT #'s EXP. DATES 07-12283-00 03/26/2013 07-12284-00 03/27/2013 07-12285-00 03/28/2013 07-12289-00 04/01/2013 07-12290-00 04/02/2013 07-12292-00 04/04/2013 07-12293-00 04/05/2013 07-12296-00 04/08/2013 07-12297-00 04/09/2013 07-12298-00 04/10/2013 07-12300-00 04/12/2013 07-12303-00 04/15/2013 07-12305-00 04/17/2013 07-12306-00 04/18/2013 07-12307-00 04/19/2013 07-12310-00 04/22/2013 07-12312-00 04/24/2013 07-12313-00 04/25/2013 07-12314-00 04/26/2013 07-12317-00 04/29/2013 07-12320-00 05/02/2013 07-12321-00 05/03/2013 07-12324-00 05/06/2013 07-12325-00 05/07/2013 07-12327-00 05/09/2013 07-12331-00 05/13/2013 07-12333-00 05/15/2013 07-12335-00 05/17/2013 07-12338-00 05/20/2013 07-12340-00 05/22/2013 07-12341-00 05/23/2013 07-12345-00 05/27/2013 07-12346-00 05/28/2013 07-12347-00 05/29/2013 07-12349-00 05/31/2013 07-12352-00 06/03/2013 07-12354-00 06/05/2013 07-12355-00 06/06/2013 07-12362-00 06/13/2013 07-12363-00 06/14/2013 07-13002-00 06/19/2013 07-13003-00 06/20/2013 07-13004-00 06/21/2013 07-13007-00 06/24/2013 07-13008-00 06/25/2013 07-13009-00 06/26/2013 07-13010-00 06/27/2013 07-13014-00 07/01/2013 07-13015-00 07/02/2013 07-13017-00 07/04/2013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1971-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1971-2013Class IITerminatedVoluntary: Firm initiated