Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0561-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module
- Product code
- JMG Radioimmunoassay (Two-Site Solid Phase), Ferritin
- Reason for recall
- Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent an Urgent Medical Device Recall letter dated May 2013 to all affected customers. The letter identified the product and lots, the reason for the field action, the risk to health and the action(s) to be taken. Customers were instructed to discontinue use and discard inventory of the affected product. Siemens will replace any unused inventory of the affected lots at no charge. For questions contact your Siemens Technical Center or your local Siemens technical support representative.
- Quantity in commerce
- 17,719
- Distribution
- US Nationwide and OUS.
- Product codes and lots
- Lot numbers RD23031, exp 2013-04-29; RD23111, exp 2013-05-07; RD23241, exp 2013-05-19; RD23311, exp 2013-06-04; and RD23461, exp 2013-06-11
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0561-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.