Z-1570-2013Class I

Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.)...

Medtronic Neuromodulation / initiated 2013-06-03 / Terminated / root cause: design

Medtronic Neuromodulation began a recall of Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. Product Usage: The... on . FDA classified it as Class I and gives the reason as "Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1570-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Neuromodulation
Product
Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. Product Usage: The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II infusion system include the pump, catheter, and catheter accessories. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids: Chronic intraspinal (epidural and intrathecal) infusion of preservative-free morphine sulfate sterile solution in the treatment of chronic intractable pain, Chronic intrathecal infusion of preservative-free ziconotide sterile solution for the management of severe chronic pain, Chronic intrathecal infusion of Lioresal Intrathecal (baclofen Injection ) in the management of serve spasticity, Chronic Intravascular Infusion of Floxuridine (FUDR) or methotrexate for the treatment of primary or metastatic cancer.
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
Medtronic is providing Healthcare Professionals with important safety information and patient management recommendations regarding the unintended delivery of drug during the priming bolus function for the SynchroMed implantable infusion pump. The unintended delivery of drug can contribute to patient overdose or underdose symptoms which may be clinically relevant.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic sent an Urgent Medical Device Correction letter dated May 2013, to all affected customers. The letter described the background and nature of the issue and provided recommendations for patient management and monitoring after initiation of Intrathecal Therapy. The letter directed customers to continue to use the priming bolus precedure; to monitor all patients following start or restart of intrathecal therapy; consider priming the pump prior to implant; and educate caregivers and family members to recognize the signs and sypmtoms associated with intrathecal drug therapy complications. For questions contact your local Medtronic field representative or Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity in commerce
261,109 total devices estimated implanted Worldwide
Distribution
Worldwide Distribution - USA Nationwide
Product codes and lots
all SynchroMed II pumps

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1570-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1570-2013Class ITerminatedVoluntary: Firm initiated
Z-1571-2013Class ITerminatedVoluntary: Firm initiated