Z-0433-2014Class II

syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab...

Siemens Healthcare Diagnostics Inc. / initiated 2013-05-10 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R)... on . FDA classified it as Class II and gives the reason as "Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0433-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
syngo(R) Lab Data Manager - Version VA11B and VA12A Systems Securely consolidates data from multiple lab instruments, the LIS and Siemens StreamLAB(R) Automation Solutions. It provides autoverification, integrated QC capabilities, browser-based connectivity supporting up to 25 concurrent users, remote services to help keep systems running optimally and also has audit trail capabilities.
Product code
NSX
Reason for recall
Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent an Urgent Medical Device Correction letter in May 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the letter in its entirety and ensure that all operators understand the information presented. Specific actions needed to be implemented by the customer were outlined in the letter. Customers were instructed to retain the letter with their laboratory records, and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.
Quantity in commerce
262
Distribution
Worldwide Distribution: USA (nationwide) and internationally to Canada.
Product codes and lots
Siemens Material Number 10800057 (Version VA11B) and 10803188 (Version VA12A)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0433-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0433-2014Class IITerminatedVoluntary: Firm initiated