Z-1622-2013Class II

MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and...

Maquet Cardiovascular Us Sales, LLC / initiated 2013-04-26 / Terminated / root cause: design

Maquet Cardiovascular Us Sales, Llc began a recall of MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures... on . FDA classified it as Class II and gives the reason as "MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1622-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
MAQUET Getinge Group Mobile Operating Table YUNO OTN GA 1433.01 EN 05. The YUNO OTN is used to support and position patients for surgical procedures immediately before, during and after the surgical phase. The table top is radio translucent and enables intraoperative use of X-ray equipment.
Product code
LGX Table, Examination, Medical, Powered
Reason for recall
MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,
Root cause tag
design
FDA root cause
Device Design
Action
MAQUET Getinge Group issued an Urgent Medical Device Recall letter/"Field Action Response Form dated April 26, 2013, via Federal Express to all affected customers. The recall notification identifies the product, issue with the product, and measures to be taken by the user. Customers were instructed that the firm would contact them to arrange for a loaner YUNO table and extension device while the table and extension device are upgraded. Upon completion of the upgrades, MAQUET would reinstall their upgraded YUNO table and extension device and remove the loaner device. Customers were instructed to complete and return the enclosed Field Action Response Form, acknowledging their receipt and understanding of the communication by e-mailing a scanned copy to [email protected] or via fax at 973-807-1658. Customers with questions were instructed to contact their MAQUET representative or MAQUET Technical Support at 888-627-8383, option 3, followed by option 1. For questions regarding this recall call 973-709-7691.
Quantity in commerce
13 Tables; 24 Extension devices
Distribution
Worldwide Distribution - USA including Arizona, Georgia, North Dakota, South Carolina, Texas and Tennessee, Internationally to United Arab Emirates, Austria, Brazil, Canada, Switzerland, China, Germany, Spain, France, United Kingdom, Italy, Japan, Singapore, Thailand, Turkey, Serbia, and South Africa.
Product codes and lots
YUNO OTN Mobile Operating Table: 1433.01F0 Extension Device: 1433.62A1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1622-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1622-2013Class IITerminatedVoluntary: Firm initiated