Z-1421-2013Class II

Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and...

Alphatec Spine, Inc. / initiated 2013-05-06 / Terminated / root cause: design

Alphatec Spine, Inc. began a recall of Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments on . FDA classified it as Class II and gives the reason as "Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1421-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Alphatec Spine, Inc.
Product
Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve
Root cause tag
design
FDA root cause
Device Design
Action
Alphatec Spine initiated recall by contacting customers via telephone and/or email on 05/06/2013. An Urgent Medical Device Recall Notification letter along with a response form were sent via FedEx to all affected accounts. The letter identified the product, problem, and actions to be taken by the customers.
Quantity in commerce
1,814
Distribution
Nationwide Distribution.
Product codes and lots
83010 85001 85002 83005-30  83005-35  83005-40  83005-45  83005-50  83006-30  83006-35  83006-40  83006-45  83006-50  83006-55  83006-60  83007-30  83007-35  83007-40  83007-45  83007-50  83007-55  83008-30  83008-35  83008-40  83008-45  83008-50  83008-55  83008-80  82001-1    Catalog Numbers: LA-HSS0010  LA-HDOO01  LA-CTOO02  LA-PS0530  LA-PS0535  LA-PS0540  LA-PS0545  LA-PS0550  LA-PS0630  LA-PS0635  LA-PS0640  LA-PS0645  LA-PS0650  LA-PS0655  LA-PS0660  LA-PS0730  LA-PS0735  LA-PS0740  LA-PS0745  LA-PS0750  LA-PS0755  LA-PS0830  LA-PS0835  LA-PS0840  LA-PS0845  LA-PS0850  LA-PS0855  LA-PS0880  LA-SOO01   Lots Affected: All

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1421-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1421-2013Class IITerminatedVoluntary: Firm initiated