510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1542-2013Class II

QuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The...

Edwards Lifesciences LLC / initiated 2013-05-14 / Terminated / root cause: manufacturing process

Edwards Lifesciences, LLC began a recall of QuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The product is packaged sterile and non-pyrogenic in a... on . FDA classified it as Class II and gives the reason as "Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1542-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Edwards Lifesciences LLC
Product
QuickDraw Venous Cannula with Insertion Kit. ThruPort Systmes. Models QD22 (22 Fr) and QD25 (25 Fr). The product is packaged sterile and non-pyrogenic in a sealed, peel-type pouch. Two pouches (cannula and accessories) go into one shelf carton. One shelf carton goes in one shipper carton. Use of the QuickDraw venous cannula is indicated for patients undergoing cardiopulmonary bypass. The QuickDraw venous cannula serves to drain non-oxygenated blood from the venae cavae or right atrium during cardiopulmonary bypass. The product is packaged sterile and nona pyrogenic in a sealed, peel-type pouch.
Product code
DWS Instruments, Surgical, Cardiovascular
Reason for recall
Edwards is recalling the QuickDraw Venous Cannula due to complaints of cannula separation during withdrawal of the device when the Percutaneous insertion method is used.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm, Edwards Lifesciences, sent an "URGENT FIELD SAFETY NOTICE PRODUCT RECALL - ACTION REQUIRED" notice dated May 14, 2013 to its customers. The letter described the product, problem, and actions to be taken. The customers were instructed to review our entire inventory for any QuickDraw Venus Cannula that has not expired; quarantine affected product from your inventory; discontinue use and return unused devices; complete and return the acknowledgment form via fax to Edwards Customer Service on 800-422-9329 within three days of receipt of this Field Safey Notice; contact Customer Service at 800-424-3278 to obtain a Returned Goods Authorization number and replacement product; return product to Edwards Lifesciences, Attn: Santosh Bhagat, 12050 Lone Peak Drive, Draper, UT 84020 Attention: RECALL, RGA#xxxxxx. If you have any questions that have not been answered by this notice, please call Edwards Customer Service at 800-424-3278 from the hours of 6:00AM - 4:30PM PST.
Quantity in commerce
22,616
Distribution
Worldwide distribution: US Nationwide and country of Canada.
Product codes and lots
Model QD22: Lot No. 59049758, 59054488, 59067502, 59074170, 59081626, 59083736, 59091652, 59094887, 59114755, 59117451, 59134508, 59149651, 59154310, 59178579, 59201243, 59212118, 59222657, 59233636, 59233637, 59246684, 59246701, 59255899, 59255900, 59269439, 59284780, 59287725, 59296462, 59299605, 59315339, 59320575, 59334652, 59356129, 59360072, 59365198, 59365199, 59374009, 59384309, 59390552, 59394166, 59394167, 59421250, 59448841  Model QD25: Lot No. 59046082,59046847,59052393,59052395,59056270,59059029,59060798,59062681,59069772,59073141,59077234,59079573,59085168,59085169,59097289,59101637,59106005,59110720,59122491,59126281,59131923,59140133,59142266,59143453,59145027,59158524,59158579,59163164,59184252,59191107,59201246,59207044,59209886,59214360,59214361,59216991,59222658,59222659,59233651,59246685,59246703,59255901,59268625,59284781,59287728,59296463,59299610,59313057,59320582,59320583,59334656,59344191,59346468,59352608,59356135,59360084,59374014,59374015,59414791,59425355,59437699,59437700,59448853

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1542-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1542-2013Class IITerminatedVoluntary: Firm initiated