Z-1563-2013Class II

Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and...

Microgenics Corporation / initiated 2013-05-13 / Terminated

Microgenics Corporation began a recall of Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS... on . FDA classified it as Class II and gives the reason as "It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1563-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Microgenics Corporation
Product
Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS Bilirubin is intended for use as a consistent test sample of known concentration for monitoring assay conditions in many clinical laboratory determinations. Includes Bilirubin with patient serum specimens when assaying for any of the listed constituents. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument.
Product code
JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
Reason for recall
It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability.
FDA root cause
Under Investigation by firm
Action
ThermoFisher Scientific sent a Product Recall Notification letter dated May 13, 2013, by Certified US mail to all affected customer. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed 1. Determine if you are using or have inventory of the recalled product. 2. Discontinue use of the product immediately. 3. Evaluate quality controls. They may vary among laboratories and should be evaluated according to laboratory procedures. 4. Retain a copy of the letter for laboratory records. 5. If any of the affected lots of MAS Bilirubin had been forwarded to another laboratory, please provide a copy of the recall letter to them. 6. Complete and return the response form within 5 days. Letters also provide an alternative product - MAS Bilirubin part number BC103. Questions or concerns may be directed to Technical Services at 1-800-232-3342 or 510-979-5417.
Quantity in commerce
188
Distribution
Nationwide Distribution: US including the states of WI, UT, FL NY, PA, IL, CO, AZ and NJ.
Product codes and lots
Catalog Number: 10014432;  Lot Numbers:  1MQ310, Expiration Date: 31 Dec 2013, 2CQ310, Expiration Date: 31 Mar 2014, 2KQ310, Expiration Date: 31 Oct 2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1563-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1563-2013Class IITerminatedVoluntary: Firm initiated