Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and...
Microgenics Corporation began a recall of Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS... on . FDA classified it as Class II and gives the reason as "It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1563-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microgenics Corporation
- Product
- Thermo Scientific MAS Bilirubin for Siemens and Dimension Vista Systems, , Product is manufactured and distributed by Microgenics Corporation, Fremont, CA MAS Bilirubin is intended for use as a consistent test sample of known concentration for monitoring assay conditions in many clinical laboratory determinations. Includes Bilirubin with patient serum specimens when assaying for any of the listed constituents. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and instrument.
- Product code
- JJX Single (Specified) Analyte Controls (Assayed And Unassayed)
- Reason for recall
- It has been confirmed from in-house testing that Direct Bilirubin (DBIL), and Total Bilirubin (TBIL) claimed in Thermo Scientific MAS Bilirubin Systems are displaying reduced stability.
- FDA root cause
- Under Investigation by firm
- Action
- ThermoFisher Scientific sent a Product Recall Notification letter dated May 13, 2013, by Certified US mail to all affected customer. The letter identified the product the problem and the action needed to be taken by the customers. Customers were instructed 1. Determine if you are using or have inventory of the recalled product. 2. Discontinue use of the product immediately. 3. Evaluate quality controls. They may vary among laboratories and should be evaluated according to laboratory procedures. 4. Retain a copy of the letter for laboratory records. 5. If any of the affected lots of MAS Bilirubin had been forwarded to another laboratory, please provide a copy of the recall letter to them. 6. Complete and return the response form within 5 days. Letters also provide an alternative product - MAS Bilirubin part number BC103. Questions or concerns may be directed to Technical Services at 1-800-232-3342 or 510-979-5417.
- Quantity in commerce
- 188
- Distribution
- Nationwide Distribution: US including the states of WI, UT, FL NY, PA, IL, CO, AZ and NJ.
- Product codes and lots
- Catalog Number: 10014432; Lot Numbers: 1MQ310, Expiration Date: 31 Dec 2013, 2CQ310, Expiration Date: 31 Mar 2014, 2KQ310, Expiration Date: 31 Oct 2014
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1563-2013 | Class II | Terminated | Voluntary: Firm initiated |