8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use...
Intuitive Surgical, Inc. began a recall of 8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments on . FDA classified it as Class II and gives the reason as "Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1431-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- 8mm Precise Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Intuitive Surgical sent an "Urgent Medical Device Correction" letter dated May 8, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Customer Service at 800-876-1310 for questions regarding this notice.
- Quantity in commerce
- 697 units - all products
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.
- Product codes and lots
- Model number: 400110-10; Lot numbers: M10120807 M10120817 M10120831 M10120910 M10121008 M10121012 M10121024 M10121115 M10121123 M10121130 M10121206 M10121210 M10130124 M10130131 M10130205 M10130304 M10130319 M10130321
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1431-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1432-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1433-2013 | Class II | Terminated | Voluntary: Firm initiated |