Z-1432-2013Class II

8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple...

Intuitive Surgical, Inc. / initiated 2013-05-08 / Terminated / root cause: design

Intuitive Surgical, Inc. began a recall of 8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments on . FDA classified it as Class II and gives the reason as "Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1432-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Intuitive Surgical, Inc.
Product
8mm Maryland Bipolar Forceps used in conjunction with the da Vinci Surgical System, Model IS1200. Multiple use electrosurgical endoscopic instruments.
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Reports of dislodging of the Cautery Plug Insert from the Cautery Plug Riser, preventing a bipolar cautery cable from being properly connected which results in loss of cautery function of Bipolar Instruments.
Root cause tag
design
FDA root cause
Device Design
Action
Intuitive Surgical sent an "Urgent Medical Device Correction" letter dated May 8, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Customer Service at 800-876-1310 for questions regarding this notice.
Quantity in commerce
697 - all products
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, CA, CO, FL, IL, KY, MI, MS, NJ, NY, PA, TX, UT, VA, WA, and WI, and the countries of Austria, Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Netherlands, Norway, Qatar, Romania, Spain, Sweden, Switzerland, Taiwan, United Kingdom, Saudi Arabia, Singapore, and Venezuela.
Product codes and lots
Model number: 400172-08; Lot numbers: M10120815 M10120817 M10120831 M10120913 M10120918 M10121008 M10121024 M10121120 M10121123 M10121203 M10121210 M10121215 M10130220 M10130222 M10130311 M10130315 M10130327 M10130411.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1432-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1431-2013Class IITerminatedVoluntary: Firm initiated
Z-1432-2013Class IITerminatedVoluntary: Firm initiated
Z-1433-2013Class IITerminatedVoluntary: Firm initiated