Z-1509-2013Class II

FLEXSTENT Biliary Self-Expanding Stent System

Cordis US Corp. / initiated 2013-04-23 / Terminated / root cause: packaging

Cordis Corporation began a recall of FLEXSTENT Biliary Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Biliary SES product is indicated for use in the palliation of malignant... on . FDA classified it as Class II and gives the reason as "Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1509-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
FLEXSTENT Biliary Self-Expanding Stent System. Product Usage: The FSS FLEXSTENT Biliary SES product is indicated for use in the palliation of malignant strictures in the biliary tree.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
Holes and/or tears can be introduced in the FLEXSTENT inner pouch under some conditions, and to a lesser degree, in the outer pouch.
Root cause tag
packaging
FDA root cause
Package design/selection
Action
Cordis sent an Urgent Medical Device Correction and Potential Removal letter dated April 24, 2013 e-mail to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately identify and set aside all affected products to prevent further use of the affected product, until inspected. Use only product identified with a blue dot next to the lot number on the carton face label, carton end label, and outer pouch label, or product with upgraded sterile barrier system (not yet available, as of the date of this letter), review, complete, sign and return the enclosed Acknowledgement Form and maintain a copy of this notice with the affected product. For questions related to the issue, please contact our FSS/Cordis representative, or please contact Cordis Direct at (800) 781-0282, Monday through Friday from 9:00 AM to 5:00 PM EDT.
Quantity in commerce
1357 units
Distribution
Worldwide Distribution - USA (nationwide) and the states of: AZ, CA, CT, FL, GA, IA, IL, LA, MD, MI, MS, NC, NJ, NY, OH, PA, RI, SC, TX, WA, WI, WV, and the countries of: Czech Republic, Italy, Netherlands, and New Zealand
Product codes and lots
27 Catalog Numbers   FLX-50030-B2, FLX-60030-B2, FLX-70030-B2, FLX-80030-B2, FLX-50040-B2, FLX-60040-B2, FLX-70040-B2, FLX-80040-B2, FLX-50060-B2, FLX-60060-B2, FLX-70060-B2, FLX-80060-B2, FLX-50080-B2, FLX-60080-B2, FLX-70080-B2, FLX-80080-B2, FLX-50100-B2, FLX-60100-B2, FLX-70100-B2, FLX-80100-B2, FLX-50120-B2, FLX-60120-B2, FLX-70120-B2, FLX-80120-B2, FLX-60150-B2, FLX-70150-B2, FLX-80150-B2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1509-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1507-2013Class IITerminatedVoluntary: Firm initiated
Z-1508-2013Class IITerminatedVoluntary: Firm initiated
Z-1509-2013Class IITerminatedVoluntary: Firm initiated
Z-1510-2013Class IITerminatedVoluntary: Firm initiated
Z-1511-2013Class IITerminatedVoluntary: Firm initiated