Z-1521-2013Class II

Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for...

Applied Medical Resources Corp / initiated 2013-05-01 / Terminated / root cause: manufacturing process

Applied Medical Resources Corp began a recall of Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or... on . FDA classified it as Class II and gives the reason as "Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1521-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Applied Medical Resources Corp
Product
Applied Medicals Epix and Direct Drive Laparoscopic Graspers, Model Numbers C4130 and C4140. Used for grasping and manipulating tissue during general or laparoscopic surgery.
Product code
NWV Laparoscopic Accessories, Gynecologic
Reason for recall
Applied Medical is conducting a voluntary recall on specific lot numbers of its Epix and Direct Drive Laparoscopic Graspers. When the ratchet trigger is activated, if a great enough force is applied, there is a potential for the trigger to fracture and the jaws to remain in the closed position. The likelihood of this situation to occur and result in permanent patient injury is highly unlikely;
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Applied Medical notified customers of the recall beginning on May 13, 2013. The "URGENT: MEDICAL DEVICE RECALL" letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact the firm at (949) 713-8041.
Quantity in commerce
127,305 units
Distribution
Worldwide Distribution: USA Nationwide and countries of: Argentina, Austria, Australia, Bahrain, Belgium, Canada, Chile, Columbia, Denmark, Ecuador, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Jordan, Kuwait, Malaysia, Lebanon, Lybia, Mexico, Netherlands, New Zealand, Pakistan, Peru, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Spain, South Africa, Switzerland, Taiwan-Province of China,Turkey, United Arabs Emirates United Kingdom, Uruguay, and Venezula-Bolivarian Republic.
Product codes and lots
Model Numbers C4130 and C4140.  MODEL NUMBERS    C4130- 5mm x 35cm Epix/Direct Drive Laparoscopic Grasper  LOT NUMBERS  1145332, 1145943, 1145944, 1145945, 1145992, 1145993, 1145994, 1146371, 1147900, 1147901, 1147955, 1148172, 1148173, 1148174, 1148180, 1148181, 1148184, 1148190, 1148191, 1148192, 1148194, 1148201, 1148203, 1148677, 1148765, 1148766, 1148767, 1148835, 1149040, 1149041, 1149690, 1149757, 1150168, 1150169, 1150170, 1150490, 1150634, 1151246, 1151247, 1151248, 1152274, 1152728, 1153332, 1154394, 1154637,  1155135, 1155473, 1155948, 1156004, 1156010, 1156011, 1156012, 1156270, 1156360, 1156452, 1156577, 1156857, 1157218, 1157679, 1157950, 1157951, 1157952, 1157953, 1157954, 1158052, 1158131, 1158216, 1158217, 1158425, 1159491, 1160167, 1161136, 1161137, 1161398, 1161638, 1162112, 1162503, 1162504, 1162588, 1163399, 1163558, 1163559, 1163575, 1164580, 1164581, 1164582, 1164583, 1168124, 1168637, 1168638, 1169117, 1171168, 1171631, 1172170, 1172668, 1173153, 1174995, 1174996, 1176616, 1177407, 1177408, 1178109, 1180295, 1181554, 1182940, 1183774, 1184416, 1184959, 1187241, 1188110, 1188745, 1189773, 1191368, 1191809.  C4140- 5mm x 45cm Epix/Direct Drive Laparoscopic  Grasper  1145559, 1147851, 1147852, 1147960, 1148210, 1148211, 1149038, 1149039, 1149689, 1149936, 1150488, 1150489, 1151249, 1152594, 1153330, 1155645, 1157217, 1157949, 1158644, 1160033, 1160170, 1162111, 1167871, 1169855, 1171497, 1172009, 1173587, 1174121, 1177578, 1179347, 1180753, 1181346, 1182924, 1183769, 1184961, 1188111, 1189182  Various Kits Containing Either C4130 or C4140 1148079, 1148139, 1148145, 1148250, 1148514, 1148770, 1148780, 1149100, 1149605, 1150455, 1151831, 1153074, 1153702, 1154931, 1154997, 1155332, 1156581, 1156812, 1157090, 1158535, 1160152, 1160227, 1160388, 1161766, 1162189, 1162343, 1163348, 1163881, 1164590, 1164796, 1164831, 1165443, 1166520, 1168227, 1169190, 1169456, 1169620, 1169772, 1170151, 1170378, 1171280, 1171878, 1171879, 1172330, 1173319, 1173903, 1174093, 1175987, 1175991, 1176171, 1178210, 1178498, 1178581, 1179376, 1182679, 1182972, 1183303, 1184673, 1185599, 1185762, 1186257, 1186868, 1187082, 1188827, 1189581, 1190574, 1191055, 1193112, 1193147, 1193648, 1193688, 1194964,  1194965

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1521-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1521-2013Class IITerminatedVoluntary: Firm initiated