Z-1484-2013Class II

Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to...

Zimmer Surgical SA / initiated 2013-05-07 / Terminated / root cause: manufacturing process

Zimmer Surgical Inc began a recall of Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities on . FDA classified it as Class II and gives the reason as "This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1484-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer Surgical SA
Product
Zimmer Air Dermatome II, REF 00-8851-001-00 handpiece Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities
Product code
GFD Dermatome
Reason for recall
This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm, Zimmer, sent an "URGENT: DEVICE REMOVAL" notice dated May 7, 2013, to all consignees/customers via FED Ex and Distributors via E-mail. Consignees who had devices repaired received telephone calls as well as the mailed Recall Notice. The Notice identified the devices being recalled; possible adverse events and directions to immediately stop use of and remove affected product, and obtain an RGA (Return Goods Authorization) number by calling (800)348-2759 or email information to [email protected] . to return the devices for repair to:Zimmer Surgical, Attn: QA/RA Dept.- Recall, 200 West Ohio Avenue Dover, Ohio 44622 USA; and to complete and return the Response Form-Immediate Response Required via email to [email protected]. If you have further distributed affected product(s), please provide the customer's information on the Business Response Form to Zimmer. If you have any questions regarding this action, please email us at CorporateQuality. [email protected] or call 1-330-364-0989.
Quantity in commerce
126 units
Distribution
Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.
Product codes and lots
Serial Number 500191 , 500192,500193, 500194,500196,500197,500198,500199, 500200, 500201,500202,500203, 500204, 500205,500206, 500207,500208,500209,500210, 500211,500212, 500213,500215,500216,500217, 500219,500220, 500221,500222, 500223, 500224, 500225,500226,500227,500228,500229,500230, 500231,500232, 500233,500234,500235,500236, 500237, 500238,500239,500240,500241 , 500242, 500243,500244,500245,500246,500247,500248,500249, 500250, 500251,500252,500253,500254, 500255,500256,500257,500258,500259, 500260,500261,500262,500263,500264, 500265,500266,500267,500268,500269, 500270,500291, 500292,500293,500295,500296,500297,500298, 500299,500300,500301 , 500302,500303, 500304, 500306, 500307,500308,500309,500310, 500331 , 500332, 500334, 500335, 500336, 500337, 500338, 500339, 500340, 500341,500342, 500343, 500344, 500345, 500346, 500347, 500348,500349, 500350, 500351 , 500352, 500353, 500354,500355, 500357, 500358, 500359, 500368, 500369,500370,500371

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1484-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1483-2013Class IITerminatedVoluntary: Firm initiated
Z-1484-2013Class IITerminatedVoluntary: Firm initiated
Z-1485-2013Class IITerminatedVoluntary: Firm initiated