Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1631-2013Class II

- Acorn Bur, Long, For Use With TPS Universal Drill, Long "AM" Attachment Head: 7.5mm Dia. x 9.5mm Length...

Medline ReNewal / initiated 2013-05-03 / Terminated / root cause: packaging

Surgical Instrument Service And Savings, Inc. began a recall of - Acorn Bur, Long, For Use With TPS Universal Drill, Long "AM" Attachment Head: 7.5mm Dia. x 9.5mm Length Ref. Lot 122665 - Cannulated Drill Bit... on . FDA classified it as Class II and gives the reason as "The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1631-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medline ReNewal
Product
- Acorn Bur, Long, For Use With TPS Universal Drill, Long "AM" Attachment Head: 7.5mm Dia. x 9.5mm Length Ref. 5130-30-75 Lot 122665 - Cannulated Drill Bit, for 3.0mm Cannulated Screws, 1.2mm Cannulation, Quick Coupling 2.0mm x 150mm Ref. 310.221 Lot 122289 - Drill Bit, For Modular Hand System, Quick Coupling 2.8mm x 165mm Ref. 310.288 Lot 122322 - Drill Bit, Quick Coupling (Gold) 2.5mm Ref. DB-2.5 Lot 122289, 122322 - Egg Micro Bur, 10 Flute, Stainless Steel 4.0mm Ref. 1608-2-35 Lot 122429, 122347 - Hall Micro Bur, Oval, Long, 8-Flute, Carbide 4.0mm x 8.0mm Ref. 5092-136 Lot 123353 - Micro Bur, Round, Stainless Steel 3.2mm Ref. 1608-2-11 Lot 122447, 122347 - Oval Bur, Fast Cutting, 8-Flutes, Carbide, 44.5mm Total Length, 18.6mm Exposure 4.0mm x 7.9mm Ref. 227-10-62 Lot 122411 - Oval Bur, Solid, 8-Flute, Medium, Head Length: 7.7mm, Carbide 4.0mm x 48mm Ref. 5300-10-901 Lot 122647 - Oval Cutting Carbide Bur, Medium, (8-Flute) 4.0mm Ref. 5091-136 Lot 123353, 122454 - Round Bur, Medium, Solid 8-Flute, Carbide 4.0mm x 48.3mm Ref. 5300-10-511 Lot 122665 - Round Bur, Micro, 8-Flute, Exposure: 28.2mm Stainless Steel 2.4mm x 54mm Ref. 1608-2-9 Lot 122447 - Twist Drill, Quick Coupling 110mm 2.5mm Ref. 310.25 Lot 122289, 122322 - Wire Pass Drill Bit 1.5mm Ref. 1608-2-59 Lot 122511 Surgical instrument motors and accessories/attachments
Product code
GFF Bur, Surgical, General & Plastic Surgery
Reason for recall
The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
MEDISISS sent a Correction Immediate Action Request letter dated May 2, 2013, to all affected customes. The letter identified the product, the problem, and the action to be taken by the customer. The letter directed the customer to inspect the products prior to use for seal integrity as specified on the product labeling: "Sterile unless package has been damaged or opened. Do not use if package is damaged." This will be followed up by a visit from the customer's sales representatives who are tasked with inspecting all the lots that may be remaining in the customer's inventory. The sales representative is directed to complete a "Verification of Seal Inspection" form and return it to MEDISISS Customer Relations. Devices with unacceptable seals are to be returned to MEDISISS. The sales representative is responsible for obtaining a shipping label from customer service to enable the devices to be returned to the recalling firm via Fed Ex. Customers with questions were instructed to call 866-866-7477.
Quantity in commerce
30
Distribution
Nationwide Distribution.
Product codes and lots
122347, 122454, 122411, 122289, 123353, 122429, 122322, 122447, 122665, 122647, 122511.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1631-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1623-2013Class IITerminatedVoluntary: Firm initiated
Z-1624-2013Class IITerminatedVoluntary: Firm initiated
Z-1625-2013Class IITerminatedVoluntary: Firm initiated
Z-1626-2013Class IITerminatedVoluntary: Firm initiated
Z-1627-2013Class IITerminatedVoluntary: Firm initiated
Z-1628-2013Class IITerminatedVoluntary: Firm initiated
Z-1629-2013Class IITerminatedVoluntary: Firm initiated
Z-1630-2013Class IITerminatedVoluntary: Firm initiated
Z-1631-2013Class IITerminatedVoluntary: Firm initiated
Z-1632-2013Class IITerminatedVoluntary: Firm initiated
Z-1633-2013Class IITerminatedVoluntary: Firm initiated
Z-1634-2013Class IITerminatedVoluntary: Firm initiated
Z-1635-2013Class IITerminatedVoluntary: Firm initiated
Z-1636-2013Class IITerminatedVoluntary: Firm initiated
Z-1637-2013Class IITerminatedVoluntary: Firm initiated
Z-1638-2013Class IITerminatedVoluntary: Firm initiated
Z-1639-2013Class IITerminatedVoluntary: Firm initiated
Z-1640-2013Class IITerminatedVoluntary: Firm initiated
Z-1641-2013Class IITerminatedVoluntary: Firm initiated
Z-1642-2013Class IITerminatedVoluntary: Firm initiated
Z-1643-2013Class IITerminatedVoluntary: Firm initiated
Z-1644-2013Class IITerminatedVoluntary: Firm initiated