Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the...
Zimmer, Inc. began a recall of Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile . Used in retracting the humeral shaft posteriorly and exposing the entire... on . FDA classified it as Class II and gives the reason as "This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1620-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Bigliani/Flatow The Complete Solution Fukuda Retractors Large Nonsterile 4305-19-40. Used in retracting the humeral shaft posteriorly and exposing the entire glenoid surface.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- This is a recall expansion from the 2010 recall which now includes seven additional lots that were not previously included. Zimmer, Inc. is initiating a recall of small and large Fukuda-type retractors due to fractures occurring at the grooves around the perimeter of the blades while being used during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- In 2010, Zimmer initiated a recall of the Bigliani/Flatow Fukuda Retractors due to fractures occurring at the grooves around the perimeter of the blades. Zimmer is expanding this recall to include seven lots that were not previously included. On 5/23/13, an Urgent Medical Device Recall notification letter went to all consignees explaining the recall expansion and risk associated with the recall. All consignees are to locate and quarantine the device immediately. Please return the product along with the completed inventory certification form to Zimmer. If you have any questions regarding this recall, call Zimmer at 877-946-2761.
- Quantity in commerce
- 722 units (includes 2010 market withdrawal)
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of NY, MI, MN, OH, IN, IL, VA, MD, PA, WA, KY, CT, RI, NH, TX, MT, and CA and the countries of Chile, China, South Korea, and Australia.
- Product codes and lots
- Item Number: 00-4305-019-40 with Lot Numbers 61588143 and 61612276. 2010 Market Withdrawal: Item #:430501940 Lot #'s: 60768814, 60806551, 60893241, 60946820, 60999665, 61031668, 61110808, 61170678, 61205673, 61244194, 61392456, 61430127
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1620-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1621-2013 | Class II | Terminated | Voluntary: Firm initiated |