Stryker Instruments0400-720-000 T4 Pullover Toga (L/XL) Sterile, EO; single use
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Instruments0400-720-000 T4 Pullover Toga (L/XL) Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient... on . FDA classified it as Class II and gives the reason as "Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1490-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Instruments0400-720-000 T4 Pullover Toga (L/XL) Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
- Product code
- FYA Gown, Surgical
- Reason for recall
- Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Stryker sent an Urgent Urgent Recall noitifications letter dated May 7, 2013 via certified US mail and E-mail l to all affected customers, distributors and direct accounts. The letter identified the affected products, problem, risk to health, and actions to be taken. Customers were instructed to check inventory, immediately quarantine any affected product, and open each package before discarding to prevent use. Customers were directed to fill out and return a Business Reply Form to Stericycle, Phone: 888-943-2396 Fax: 888-943-4192, Email: [email protected] verifying directions and confirming recalled product was destroyed. For questions call 269-389-4354 or 269-389-4518.
- Quantity in commerce
- 48,040 units, ( 4804 boxes)
- Distribution
- USA Nationwide Distribution
- Product codes and lots
- 1005105, 1010244, 1101004, 1104289, 1107568, 1005119, 1010261, 1101030, 1104324, 1108625, 1006129, 1011276, 1101047, 1105390, 1108672, 1006140, 1011296, 1102073, 1105355, 1109730, 1007151, 1012314, 1102127, 1106431, 1109782, 1007163, 1012337, 1102131, 1106467, 1109858, 1009213, 1012346, 1104253, 1107515.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1489-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1490-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1491-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1492-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1493-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1494-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1495-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1496-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1497-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1498-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1499-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1500-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1501-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1502-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1503-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1504-2013 | Class II | Terminated | Voluntary: Firm initiated |