Z-1968-2013Class II

Terumo Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion...

Terumo Cardiovascular Systems Corporation / initiated 2013-07-26 / Terminated / root cause: software

Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal... on . FDA classified it as Class II and gives the reason as "The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1968-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1: 100/120V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
Root cause tag
software
FDA root cause
Software design
Action
Terumo CVS, sent an URGENT MEDICAL DEVICE RECALL notification letter dated July 26, 2013, to all affected customers explaining the correction to the operators manual, potential hazards and what to do in response to the issue. If you have any questions or concern please contact Terumo CVS Customer Service 1-800-521-2818 Recall Fax 1-734-741-6149 Customer Service Hours: Monday Friday, 8 AM - 6 PM ET
Quantity in commerce
1657
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Australia, Dubai, Hong Kong, Indonesia, Singapore, China, Mexico, British Columbia, Taiwan, Thailand, Chile, Malaysia, Japan, Belgium, United Arab Emirates (UAE), South Korea, Manila and Canada.
Product codes and lots
Catalog number: 801763

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1968-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1968-2013Class IITerminatedVoluntary: Firm initiated
Z-1969-2013Class IITerminatedVoluntary: Firm initiated