Terumo Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion...
Terumo Cardiovascular Systems Corporation began a recall of Terumo Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal... on . FDA classified it as Class II and gives the reason as "The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1969-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Terumo Advanced Perfusion System 1: 220/240V AC, Advanced Perfusion System 1 The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns " or similar equipment.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Terumo CVS, sent an URGENT MEDICAL DEVICE RECALL notification letter dated July 26, 2013, to all affected customers explaining the correction to the operators manual, potential hazards and what to do in response to the issue. If you have any questions or concern please contact Terumo CVS Customer Service 1-800-521-2818 Recall Fax 1-734-741-6149 Customer Service Hours: Monday Friday, 8 AM - 6 PM ET
- Quantity in commerce
- 1657
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries of Australia, Dubai, Hong Kong, Indonesia, Singapore, China, Mexico, British Columbia, Taiwan, Thailand, Chile, Malaysia, Japan, Belgium, United Arab Emirates (UAE), South Korea, Manila and Canada.
- Product codes and lots
- Catalog number: 801764
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1968-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1969-2013 | Class II | Terminated | Voluntary: Firm initiated |