Z-1877-2013Class II

Carestream DRX Evolution System, X-Ray System --- The DRX-Evolution system is a permanently installed...

Eastman Kodak Company / initiated 2013-06-03 / Terminated

Carestream Health Inc. began a recall of Carestream DRX Evolution System, X-Ray System --- The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for... on . FDA classified it as Class II and gives the reason as "Due to potential for an x-ray exposure technique change". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1877-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Carestream DRX Evolution System, X-Ray System --- The DRX-Evolution system is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Due to potential for an x-ray exposure technique change.
FDA root cause
Radiation Control for Health and Safety Act
Action
Carestream will make the correction which includes the installation of a SW patch to the affected customer sites, and to any other affected customer sites utilizing the DRX Evolution with V5. 5 SW outside of the US. --- Once receiving FDA approval on the corrective action plan, the correction will be implemented to all affected devices at identified customer sites: 1. A notification letter was sent to all customers identifying the issue and highlighting the approved procedures of the Corrective Action Plan. The letters were sent on 8/16/13 via Certified Mail, Return Receipt Requested. 2. Carestream Health will contact the customer to schedule a service visit to install software updates to correct this issue. The installation time will be approximately one hour. 3. As stated in the customer notification letter, this software update will be provided to the customer at no expense. 4. All software updates are estimated to be completed within 1-3 months. 5. Carestream Health will provide monthly progress reports on the effectiveness of the plan, including the number of units corrected. Further questions please call (800) 328-2910.
Quantity in commerce
US: 49 units; Foreign: 183 units (revised)
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Austria, Australia, Belgium, Canada, China, Finland, France, Germany, Italy, Korea, Kuwait, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Product codes and lots
Carestream DRX Evolution Systems using V5.5.410.33 MR2 Software

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1877-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1877-2013Class IITerminatedVoluntary: Firm initiated