Soma v Automated Breast Ultrasound System (ABUS), Automated Breast Ultrasound. Indicated as an adjunct to...
U-systems Inc began a recall of Soma v Automated Breast Ultrasound System (ABUS), Automated Breast Ultrasound. Indicated as an adjunct to mammography for breast cancer screening on . FDA classified it as Class II and gives the reason as "Devices labeled for an intended use not included in the existing 510(k) or PMA, in that, the user manual contains the Indication for Use of the soma v Platinum ABUS [a 510(k) product]". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1587-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- U-Systems, Inc.
- Product
- Soma v Automated Breast Ultrasound System (ABUS), Automated Breast Ultrasound. Indicated as an adjunct to mammography for breast cancer screening.
- Product code
- PAA Automated Breast Ultrasound
- Reason for recall
- Devices labeled for an intended use not included in the existing 510(k) or PMA, in that, the user manual contains the Indication for Use of the soma v Platinum ABUS [a 510(k) product].
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On 6/5/13, an "URGENT MEDICAL DEVICE CORRECTION" letter was distributed to all affected users, with a description of the problem and user corrective action steps. One consignee received an Urgent Notice to Customer - Medical Device Correction letter on 5/16/13. Contact the firm at 408-328-7925 for questions regarding this notice.
- Quantity in commerce
- 3 units
- Distribution
- Distributed in California and Illinois.
- Product codes and lots
- Serial numbers: H1280099, H1180094, B1380105
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1587-2013 | Class II | Terminated | Voluntary: Firm initiated |