510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1613-2013Class II

SYNCHRON System(s) Lactate (LACT) Reagent

Siemens Healthcare Diagnostics Inc. / initiated 2013-05-28 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX System(s), UniCel DxC 600/800... on . FDA classified it as Class II and gives the reason as "Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1613-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX System(s), UniCel DxC 600/800 System(s) and Synchron Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).
Product code
KHP Acid, Lactic, Enzymatic Method
Reason for recall
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
Root cause tag
manufacturing process
FDA root cause
Under Investigation by firm
Action
Beckman Coulter Inc. sent an Urgent Product Correction letter dated May 28, 2013 to affected customers. The letter identified the affected product, the issue, impact, actions to be taken, and resolution. The letter instructed customers to discontinue use of the affected product and request a replacement product. For questions contact Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support/ or call 800-854-3633 in the United States and Canada. Outside the United States or Canada, contact your local Beckman Coulter Representative.
Quantity in commerce
3,353 units
Distribution
Worldwide Distribution - USA Nationwide and country of Canada.
Product codes and lots
Part Number A95550, Lot Number M210077

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1613-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1613-2013Class IITerminatedVoluntary: Firm initiated