ORTHO Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO Summit System (OSS) is a modular...

Ortho-Clinical Diagnostics, Inc. / initiated 2013-05-06 / Terminated / root cause: design

Ortho-Clinical Diagnostics began a recall of ORTHO Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO Summit System (OSS) is a modular, integrated system for use with licensed blood screening... on . FDA classified it as Class II and gives the reason as "There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1651-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
ORTHO Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing.
Product code
MZA Test, Equipment, Automated Bloodborne Pathogen
Reason for recall
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Root cause tag
design
FDA root cause
Device Design
Action
Ortho Clinical Diagnostics (OCD) sent an Urgent Product Correction Notification letters dated May 6, 2013 and response forms to all affected customers. Ortho Clinical Diagnostics informed their customers of the issue and advised if they detect the smell of melting plastic or observe smoke originating from your VITROS System, to immediately power down the system and contact OCD for assistance. They were also advised to consider performing the backup procedure more frequently to help prevent the loss of data on their Systems. In the near future, an OCD Field Engineer will contact customers who are affected and perform a Modification (MOD) procedure that will replace their potentially affected Wire Connector(s). Foreign affiliates were informed by email on May 6, 2013 of the issue and instructed to notify their customers of the issue and actions. OCD also found that this wire connector is commonly used in many types of computers (i.e., medical devices, personal computers, etc.). On 29 May 2013, OCD expanded their recall to incllude ORTHO Assay Software (OAS) Servers. As a precaution, OCD sent notifications to encompass all ORTHO Assay Software (OAS) Servers that potentially contain this type of wire connector. The letters (Ref. CL13-154) were sent via US Priority Mail to all customers who were shipped an affected OAS Server. Ortho Clinical Diagnostics has requested that customers with any questions or in need of additional information, please call Customer Technical Services at 1-800-421-3311.
Quantity in commerce
31 units
Distribution
Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
Product codes and lots
Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1651-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1647-2013Class IITerminatedVoluntary: Firm initiated
Z-1648-2013Class IITerminatedVoluntary: Firm initiated
Z-1649-2013Class IITerminatedVoluntary: Firm initiated
Z-1650-2013Class IITerminatedVoluntary: Firm initiated
Z-1651-2013Class IITerminatedVoluntary: Firm initiated