Z-1533-2013Class II

BARD PERIPHERAL VASCULAR VACCESS PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm

Bard Peripheral Vascular Inc / initiated 2013-05-28 / Terminated / root cause: labeling and UDI

Bard Peripheral Vascular Inc began a recall of BARD PERIPHERAL VASCULAR VACCESS PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess PTA Balloon Dilatation Catheters are recommended... on . FDA classified it as Class II and gives the reason as "Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1533-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bard Peripheral Vascular Inc
Product
BARD PERIPHERAL VASCULAR VACCESS PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.
Product code
DQY Catheter, Percutaneous
Reason for recall
Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.
Root cause tag
labeling and UDI
FDA root cause
Mixed-up of materials/components
Action
BARD sent an URGENT MEDICAL DEVICE RECALL NOTIFICATION letters dated May 30, 2013 via FedEx overnight with proof of delivery notice to all affected customers. The letter identified the affected product, reason for recall, actions to be taken, and contact information. Customers were instructed to complete the enclosed Recall and Effectivenss Check Form and fax to the attention of Recall Coordinator in Customer Service at 1-800-994-6772.
Quantity in commerce
415
Distribution
US Nationwide Distribution
Product codes and lots
Product Codes and Lot Numbers, respectively:  VA8084R, 93JW0056 VA8084, 937W0090

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1533-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1533-2013Class IITerminatedVoluntary: Firm initiated