Z-1588-2013Class II

Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed...

Philips North America LLC / initiated 2013-03-21 / Terminated / root cause: software

Philips Medical Systems (Cleveland) Inc began a recall of Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images... on . FDA classified it as Class II and gives the reason as "If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1588-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Brilliance CT 64, Ingenuity CT, Ingenuity Core, and Ingenuity Core128 scanners are whole body Computed Tomography X-ray systems. Produce cross-sectional images of the body by computer reconstruction of x-ray transmission.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol
Root cause tag
software
FDA root cause
Software design
Action
Philips Medical Systems sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated Mar 21, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Additionally, a Field Service Engineer will contact each site to schedule a time to implement the Field Change Order.
Quantity in commerce
50 units
Distribution
Nationwide Distribution-including the states of AR, AZ, DE, IN, KS, LA, MA, MN, MS, MT, NJ, NY, NY, OH, OR, PA, TN, TX, VT, WV, and WY.
Product codes and lots
The following codes are affected by this recall:System Code #728231, Serial #: 9654, 91003, 95130, 95157, 95504  System Code #728326, Serial #:  32000,300003,300004,300005,300010,300011,300012,300013, 300014,300015,300018, 300027,300041,300053,300054, 300059, 300066, 300070,300071, 300073, 300074, 30077, 300078, 300079, 300080,300082,300084,300086,300091,300092, 300095, 300101, 300102,300112, 300119, 300127, 300128, 300129, 300131, 300132, 310064, 320030  System Code #728321, Serial #: 310096, 310070  System Code #728323, Serial #: 320033

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1588-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1588-2013Class IITerminatedVoluntary: Firm initiated