Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for...
ConBio, a CynoSure Company began a recall of Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for General Dermatology, Dermatologic and General... on . FDA classified it as Class II and gives the reason as "An inconsistency was identified related to electronic display identification of the optional Multilite Dye Handpieces of the RevLite SI Laser system that may confuse user". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1599-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- ConBio, a CynoSure Company
- Product
- Multilite Dye Handpieces, an optional accessory of the RevLite SI Laser. The Revlite SI laser is intended for General Dermatology, Dermatologic and General Surgical Procedures. Which includes treatments such as tattoo removal, pigmented lesion removal, removal or lightening of hair, skin resurfacing and reduction of red pigmentation in hypertropic and keloid scars The two Multilite Dye Handpieces (HPs) are optional handpieces that can be used with the Revlite SI Laser. Manufactured by: ConBio, A Cynosure Company 47733 Fremont blvd Fremont, CA 94538
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- An inconsistency was identified related to electronic display identification of the optional Multilite Dye Handpieces of the RevLite SI Laser system that may confuse user.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm, Cynosure Inc., sent a "MEDICAL DEVICE FIELD ACTIONS NOTIFICATION" letter dated May 9, 2013 to its customers via UPS next day. The letter described the product, problem and actions to be taken. The customers were instructed to immediately stop using the Revlite SI Multilite Dye Handpieces upon receipt of this letter and complete and return the enclosed Field Corrective Action Acknowledgement Form via e-mail to: [email protected] or FAX to: 978-256-6556. A Cynosure representative will be contacting customers to schedule and arrange an exchange of the affected handpieces. If additional information is required please contact VP of Regulatory Affairs and Quality Systems at 978-367-2350 or via email at [email protected].
- Quantity in commerce
- 12 (6 of model 659-1050; 6 of model 659-1100)
- Distribution
- Worldwide distribution: USA (nationwide) including states of: IL, ME, NM and VA; and countries of: Australia and Chile.
- Product codes and lots
- 659-1050-585 nm Handpiece; Manufacturer numbers: 659-1050 / C13B1141 659-1050 / D13C1151 659-1050 / D13C1157 659-1050 / D13C1158 659-1050 / D13C1160 659-1050 / D13C1154 Model 659-1100 - 650 nm Handpiece; manufacturer numbers: 659-1100 / C13B1141 659-1100 / D13C1151 659-1100 / D13C1157 659-1100 / D13C1158 659-1100 / D13C1160 659-1100 / D13D1165.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1599-2013 | Class II | Terminated | Voluntary: Firm initiated |